Lithium Ion Battery M4605A and M4607A for use with IntelliVue Patient Monitors. The monitor is used for monitoring and recording of, and to generate alarms for, multiple physiological parameters.

Publicado el 28 de mayo de 2014 · Inicio del retiro: 4 de abril de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The risk of battery failure increases with age, when a battery remains in use longer than 3 years after date of manufacture or 500 charge-discharge cycles. Such failure can result in overheating that in rare cases can cause the battery to ignite or explode.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Philips Medical Systems, Inc.
Distribuidor
Worldwide Distribution - Including US Nationwide, Canada, Czech Republic, Denmark, Netherlands, Estonia, Finland, France, Germany, Greece, Hungary, Italy, Japan, Kazakhstan, Norway, Poland, Romania, Russia, China, Slovakia, Sweden, and Turkey.
Lote
Patient Monitor Product numbers M3002A, M8102A, M8105A, M8001A, M8002A, M8003A, M8004A, M8027A, M8001AU, M8002AU, M8003AU, and M8004AU with software releases up to and including G.0.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States