Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1629-2022

BECKMAN COULTER, INC. a. COULTER DxH Diff Pack REAGENT, DIFF PK FP DXHCOULTR, Catalog #628020. b. COULTER DxH Cleaner REAGENT, CLEANER FP DXH 10LCOULTR Catalog #628023. c. DxH 500 Series Lyse LYSE, VERSA SOL DXH 500 SERIES(0.5L/KT) Catalog # B36846. d. DxH 500 Series Cleaner CLEANER, COULTER CLENZ DXH 500SERIES (0.5L/KT) Catalog # B36868

Publicado el 31 de agosto de 2022 · Inicio del retiro: 25 de mayo de 2022

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Facility Temperature excursions (from June 2021-August 2021) prior to delivery; exposure to higher temperatures may have impacted product effectiveness. All lots received between 06/01/21-9/30/21.

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Detalles del retiro

Fabricante
Mckesson Medical-Surgical Inc. Corporate Office
Distribuidor
US Nationwide
Lote
GTIN: a. 15099590233716. b. 15099590233747. c. 15099590671860. d. 15099590671877
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States