Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1632-2025

GE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system.

Publicado el 7 de mayo de 2025 · Inicio del retiro: 21 de marzo de 2025

Resumen

FDA registra el retiro Z-1632-2025 de dispositivos médicos de Clase I, publicado el 7 de mayo de 2025 e iniciado el 21 de marzo de 2025. Empresa responsable: GE Medical Systems China Co., Ltd.; Estado: En curso; Cantidad reportada: 69 units; Lote: GTIN 00840682103954, Serial Numbers: SM823080001MA, SM823080002MA, SM823080003MA, SM823080004MA, SM824210002WA, SM824210003WA, SM824210004WA, SM824230005WA, SM824230006WA, SM823440005WA, SM823440006WA, SM823440007WA, SM823440008WA, SM823050001WA, SM823050002WA, SM823050003WA, SM823050004WA, SM823050005WA, SM823060001WA, SM823060002WA, SM824030001MA, SM824030002MA, SM824030003MA, SM824030004MA, SM824030005MA, SM824030006MA, SM824030007MA, SM824030008MA, SM824030009MA, SM824030010MA, SM823120001MA, SM824320007WA, SM823010001WA, SM823430004WA, SM824050003MA, SM824050004MA, SM824050005MA, SM824050006MA, SM824050007MA, SM824050001WA, SM824050002WA, SM823230001MA, SM823230002MA, SM824040001WA, SM824450010WA, SM823410001WA, SM823410002WA, SM823420003WA, SM823480001WA, SM823170001MA, SM823170002MA, SM823170003MA, SM823170004MA, SM823170005MA, SM823170006MA, SM823280001MA, SM823280002MA, SM823280003MA, SM824430009WA, SM823370001MA, SM823370002MA, SM824410006MA, SM824410007MA, SM824410008MA, SM824440009MA, SM824440010MA, SM824440011MA, SM824440012MA, SM824440013MA. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
GE Medical Systems China Co., Ltd.
Distribuidor
Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Guatemala, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kosovo, Kuwait, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Oman, Pakistan, State of Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tonga, Trinidad Tobago, Tunisia, T®rkiye, Uganda, Ukraine, UAE, United Kingdom, Uruguay, Uzbekistan, Venezuela, and Vietnam.
Lote
GTIN 00840682103954, Serial Numbers: SM823080001MA, SM823080002MA, SM823080003MA, SM823080004MA, SM824210002WA, SM824210003WA, SM824210004WA, SM824230005WA, SM824230006WA, SM823440005WA, SM823440006WA, SM823440007WA, SM823440008WA, SM823050001WA, SM823050002WA, SM823050003WA, SM823050004WA, SM823050005WA, SM823060001WA, SM823060002WA, SM824030001MA, SM824030002MA, SM824030003MA, SM824030004MA, SM824030005MA, SM824030006MA, SM824030007MA, SM824030008MA, SM824030009MA, SM824030010MA, SM823120001MA, SM824320007WA, SM823010001WA, SM823430004WA, SM824050003MA, SM824050004MA, SM824050005MA, SM824050006MA, SM824050007MA, SM824050001WA, SM824050002WA, SM823230001MA, SM823230002MA, SM824040001WA, SM824450010WA, SM823410001WA, SM823410002WA, SM823420003WA, SM823480001WA, SM823170001MA, SM823170002MA, SM823170003MA, SM823170004MA, SM823170005MA, SM823170006MA, SM823280001MA, SM823280002MA, SM823280003MA, SM824430009WA, SM823370001MA, SM823370002MA, SM824410006MA, SM824410007MA, SM824410008MA, SM824440009MA, SM824440010MA, SM824440011MA, SM824440012MA, SM824440013MA.
Clasificación de riesgo
Class I
Área (FDA)
device
País
China

Retiros relacionados

Preguntas sobre este retiro

¿GE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1632-2025 el 7 de mayo de 2025. Peligro reportado: Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter