FDAFDA · Alimentos y medicamentos·N.º oficial Z-1648-2013
VITROS 4600 Chemistry System (VITROS 5,1 FS System family member)(product code 6802445) Product Usage: For use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods.)
Publicado el 10 de julio de 2013 · Inicio del retiro: 6 de mayo de 2013
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There is evidence that an electrical short within the systems connector has caused an accumulation of heat which has radiated back through the wiring harness causing the wires and connectors to melt.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Ortho-Clinical Diagnostics
- Distribuidor
- Worldwide Distribution - USA Nationwide including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NY, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI & WV, and countries of: Australia, Brazil, Canada, Chile, China, Colombia, India, Japan, Mexico, PR, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain
- Lote
- Serial numbers: J46000021, J46000110 through J46000177
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States