Patient Side Cart assembly, 4-arm, for use with Intuitive Surgical, Inc. da Vinci Surgical System IS 4000. The patient-side cart is positioned at the operating room table and contains four arms that are positioned over the target patient anatomy. The endoscope attaches onto any arm. The patient-side assistant attaches/detaches the endoscope and instrument intra-operatively.
Publicado el 11 de junio de 2014 · Inicio del retiro: 9 de mayo de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Torque wrenches used to tighten certain PSC bolts were out of calibration. This could lead to over- or under-tightening of bolts that may result in system faults, mechanical sound (e.g. scraping, squeaking, rattling), cracks or vibration in the system.
Qué hacer
Voluntary: Firm initiated · Other
Detalles del retiro
- Fabricante
- Intuitive Surgical, Inc.
- Distribuidor
- Medical facilities in TX and FL
- Lote
- lot 380652-20
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States