RemB Electric Universal Driver REF 6400-099-000 For use with the CORE and TPS System sRx Only The RemB Electric Universal Driver is intended for use with the Consolidated Operating Room Equipment (CORE TM) System. When used with a variety of accessories, the RemB Electric Universal Driver is intended for surgical procedures involving driving wire or pins into bone and hard tissue. This includes but is not limited to Dental, ENT (Ear, Nose, Throat), Neuro, and Endoscopic applications. This device can also be used with the Total Performance System (TPSTM).
Publicado el 24 de julio de 2013 · Inicio del retiro: 1 de julio de 2013
Resumen
FDA registra el retiro Z-1742-2013 de dispositivos médicos de Clase II, publicado el 24 de julio de 2013 e iniciado el 1 de julio de 2013. Empresa responsable: Stryker Instruments Div. of Stryker Corporation; Estado: Finalizado; Cantidad reportada: 9; Lote: 0910505713, 1029801073, 1225400823, 1314201603, 1314300713, 1314300723, 1314300733, 1314300743, 1314300753, 1314300763, 1314300773, 1314300783, 1314300793, 1314300803, 1314300813, 1314300823, 1314300833, 1314300843, 1314300853, 1314300863, 1314300873, 1314300883, 1314300893, 1314300903, 1314804033, 1314900053, 1314906453, 1314908193. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Instruments Div. of Stryker Corporation
- Distribuidor
- Worldwide distribution: US (nationwide) including states of: AZ, FL, GA, NC, NY, SC, TX and VA; and countries of: India, Netherlands and Japan.
- Lote
- 0910505713, 1029801073, 1225400823, 1314201603, 1314300713, 1314300723, 1314300733, 1314300743, 1314300753, 1314300763, 1314300773, 1314300783, 1314300793, 1314300803, 1314300813, 1314300823, 1314300833, 1314300843, 1314300853, 1314300863, 1314300873, 1314300883, 1314300893, 1314300903, 1314804033, 1314900053, 1314906453, 1314908193
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASurgiCount+ System Application, Software Version 2.0.17, Catalog Number 0694-002-090. Used for counting,27 de julio de 2022
- FDANeptune E-SEP 165mm Blade Electrode, Catalog Number 0703-165-001 - Product Usage: The Neptune E-SEP is an6 de mayo de 2020
- FDAColor Cuff Disposable Tourniquet Cuff, 18 inch X 3 inch, 1BLA (Single Bladder), 1PRT(Single Port) Quick11 de diciembre de 2019
- FDAC2 Nerve Monitor, 8 Channel, 5140-508-280, UDI 07613327277319 Product Usage: The C2 Nerve Monitor System is9 de octubre de 2019
- FDACORE Universal Driver REF 5400- 99 Rx Only This drill may also be used with theTotal Performance System24 de julio de 2013
- FDATDHistology/Cytology Classification name: Calculator/data processing module for clinical use (862.2100).24 de julio de 2013
Preguntas sobre este retiro
¿RemB Electric Universal Driver REF 6400-099-000 For use with the CORE and TPS System sRx Only The RemB Electric Universal Driver is intended for use with the Consolidated Operating Room Equipment (CORE TM) System. When used with a variety of accessories, the RemB Electric Universal Driver is intended for surgical procedures involving driving wire or pins into bone and hard tissue. This includes but is not limited to Dental, ENT (Ear, Nose, Throat), Neuro, and Endoscopic applications. This device can also be used with the Total Performance System (TPSTM). fue retirado del mercado?
Sí. FDA publicó el retiro Z-1742-2013 el 24 de julio de 2013. Peligro reportado: The safety margin values detailed in the Engineering Design were entered incorrectly into the programming software. Handpieces that have been programmed with the incorrect parameters could result in unintended activation. There is a potential for injury to the patient or user as a result of unintended forward or reverse activation. There is also the potential that the handpiece may not reach 100
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter