Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1760-2017

HYFRECATOR 2000 HANDPIECE SHEATH, REF/Part Number 7-796-19, Rx ONLY, STERILE, Single Patient Use -- CONMED CORPORATION, 525 FRENCH ROAD UTICA, NY 13502. --- There is no problem with the Hyfrecator 2000 Electrosurgical Units or with cases of P/N 7-796-19 sold in cases as P/N 7-796-19BX or P/N 7-796-19CS.

Publicado el 19 de abril de 2017 · Inicio del retiro: 28 de febrero de 2017

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

For over a year, the Accessory Packages in the Hyfrecator 2000 ESU cartons were packaged in error by the manufacturer with non-sterile P/N 7-796-19 Handpiece Sheaths in pouches that were labeled as Sterile. This could cause an end user to use a non-sterile sheath as if it were Sterile.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
ConMed Corporation
Distribuidor
Worldwide Distribution - US (nationwide) Internationally to Australia, Canada, Germany, Italy, UK, Great Britain, and Ireland.
Lote
Affected Lot Numbers: 201511234, 201512074, 201512094, 201601064, 201603284
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta