ABL90 FLEX Analyzer, Model Numbers 393-090, with software versions below 3.4MR2 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.

Publicado el 29 de abril de 2020 · Inicio del retiro: 3 de febrero de 2020

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The firm received reports of occurrences where the barcode reader misinterpreted the contents of a locally printed barcode label used for entering patient ID or accession number into the analyzer in connection with a sample measurement. This could result is patient mixup and/or lost of sample, resulting in delayed medical treatment.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Radiometer Medical ApS
Distribuidor
Worldwide distribution - US Nationwide distribution including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and the countries of Canada, Netherlands, China, Croatia, Czech Republic, Denmark, Germany, Spain, France, Hungary, India, Japan, South Korea, United Kingdom, Poland, Australia, South Africa, Singapore, Switzerland, Turkey, Norway, Austria, Italy, Sweden, Finland.
Lote
software versions below 3.4 MR2
Clasificación de riesgo
Class II
Área (FDA)
device
País
Denmark