FDAFDA · Alimentos y medicamentos·N.º oficial Z-1766-2017
The PreludeSYNC Radial Compression Device is a sterile, single use disposable device used to assist in gaining and maintaining hemostasis of the radial and ulnar artery following catheterization procedures. It has a soft wristband with a secure hook and loop fastener. The band delivers adjustable compression of the puncture site with an inflatable bulb, and a check valve for easy inflation and deflation with a syringe. A clear curved backer plate provides optimal visualization of the puncture site and ease of placement.
Publicado el 19 de abril de 2017 · Inicio del retiro: 3 de marzo de 2017
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Merit Medical Systems, Inc. announces a voluntary field action for the PreludeSYNC Radial Compression Device because particulate generated during the manufacturing process may enter the valve, preventing a complete seal of the inflation valve.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Merit Medical Systems, Inc.
- Distribuidor
- Nationwide, Canadian and other foreign consignees. No VA/govt/military consignees.
- Lote
- Catalog Number: SRB24AC, Lot Numbers: H1068590S1, H1078177, H1078180, H1083628; Catalog Number: SRB24MED, Lot Numbers: H1068587S1,H1073974, H1078179, H1078181, H1083629, H1083630; Catalog Number: SRB29AC, Lot Numbers: H1068591S1, H1078183, H1083631; Catalog Number: SRB29MED, Lot Numbers: H1068589S1, H1078184, H1078185, H1083633.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States