DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065. The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.

Publicado el 21 de mayo de 2025 · Inicio del retiro: 20 de marzo de 2025

Resumen

FDA registra el retiro Z-1784-2025 de dispositivos médicos de Clase II, publicado el 21 de mayo de 2025 e iniciado el 20 de marzo de 2025. Empresa responsable: Beckman Coulter Inc.; Estado: En curso; Lote: Catalog Number: C63519 UDI-DI code: 14987666545058 Serial Numbers: Catalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers:. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Beckman Coulter Inc.
Distribuidor
U.S: AL, CA, GA, ID, IN, LA, MA, MI, MN, MT, NC, NV, PA, PR, TN , TX, and WI O.U.S.: Australia, Belgium, Bosnia and Herzegovina, Canada, Chile, Croatia, Egypt, Ethiopia, France, Germany, Ghana, India, Italy, Korea, Lebanon, Libya, Malaysia, Netherlands, North Macedonia, Oman, Paraguay, Philippines, Poland, Puerto Rico, Qatar, Saudi Arabia, South Africa, Spain, Switzerland, and Tunisia
Lote
Catalog Number: C63519 UDI-DI code: 14987666545058 Serial Numbers: Catalog Number: C63520 UDI-DI code: 14987666545065 Serial Numbers:
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065. The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1784-2025 el 21 de mayo de 2025. Peligro reportado: Beckman Coulter is recalling their DxC 500 AU Clinical Chemistry Analyzer because a software error causes the analyzer to not run a requested calibration order in the following scenario: when a reagent blank or calibration is ordered during sample processing and then any of the components (R1 and/or R2) depletes to zero tests, the analyzer will not be able to complete the calibration request, and the calibration order will remain pending. No further calibration orders can be processed for any assays, and the instrument refuses to accept further sample processing order after the existing calibration curves are expired. Although in-process tests will be completed, this error can cause a delay in reporting subsequent test results. No further calibration orders can be processed for any assays which may cause a delay in reporting test results.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter