STERIS Fine Traction Device, one unit per box Product Usage: The Fine Traction Device is used to provide traction and enable fine adjustments to traction and orientation of the leg during orthopedic procedures.

Publicado el 31 de julio de 2013 · Inicio del retiro: 20 de mayo de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

A bearing used in the assembly of the Fine Traction Device did not perform successfully in pre-release reliability testing; a replacement bushing passed reliability testing and was substituted per a change order. The change order authorized use of the substitute bushing in assembly of the Fine Traction Accessory; however, inadvertently the new part was not implemented. During complaint evaluati

Qué hacer

Voluntary: Firm initiated · Visit

Detalles del retiro

Fabricante
Steris Corporation
Distribuidor
USA Nationwide Distribution in the states of CO, KY, LA, VA, and WI. The affected product was also distributed to Canada.
Lote
The following serial/lot numbers are affected by this recall: 0404913150, 0404913151, 0404913152, 0404913153, 0408713001, 0408713002, 0408713003, 0408713004, 0408713005, 0408713006, 0409813097, 0409813098, 0409813099, 0409813100, 0410113001, 0410113002, 0410713024, 0410713025, 0410713026, 0410713027, 0410713028, and 0410713029.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States