Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 recall. Product Usage: The Eon Wall Charger (3701) provides the capability to recharge the IPG Battery. The charger transmits RF energy through the antenna to the IPG battery to recharge it. The Eon Portable Charger (3711) provides the capability to recharge the IPG battery. The charger transmits RF energy through the antenna to the IPG battery to recharge it. The patient can be mobile during the recharging process with a portable charger. The Eon Mini Charger (3721) provides the capability to recharge the IPG battery while stimulation is either on or off. The charger transmits RF energy through the antenna to the IPG battery to recharge it.

Publicado el 31 de julio de 2013 · Inicio del retiro: 30 de noviembre de 2012

Resumen

FDA registra el retiro Z-1785-2013 de dispositivos médicos de Clase II, publicado el 31 de julio de 2013 e iniciado el 30 de noviembre de 2012. Empresa responsable: St. Jude Medical; Estado: Finalizado; Cantidad reportada: 75,663; Lote: All serial numbers manufactured. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

There is issues of warmth or heating at the implant site during charging for the Eon Mini implantable pulse generators (IPGs). In the July 2012 letter St. Jude Medical informed customers that they would be implementing design improvements to the charger to address possible increased energy dissipation when the charger is misaligned with the IPG or the IPG is implanted too near the surface of the

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
St. Jude Medical
Distribuidor
Worldwide Distribution - USA Nationwide including Puerto Rico and countries of: Austria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Luxembourg, Martinique, Mexico, Netherlands, Poland, Russian Federal, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Sweden, Switzerland, Turkey and United Kingdom.
Lote
All serial numbers manufactured
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Eon and Eon Mini charging system, model numbers Eon -3701 (Wall Charger), 3711 (Portable Charger) and Eon Mini -3721 (Charger); This is a continuation of the firm's July 2012 recall. Product Usage: The Eon Wall Charger (3701) provides the capability to recharge the IPG Battery. The charger transmits RF energy through the antenna to the IPG battery to recharge it. The Eon Portable Charger (3711) provides the capability to recharge the IPG battery. The charger transmits RF energy through the antenna to the IPG battery to recharge it. The patient can be mobile during the recharging process with a portable charger. The Eon Mini Charger (3721) provides the capability to recharge the IPG battery while stimulation is either on or off. The charger transmits RF energy through the antenna to the IPG battery to recharge it. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1785-2013 el 31 de julio de 2013. Peligro reportado: There is issues of warmth or heating at the implant site during charging for the Eon Mini implantable pulse generators (IPGs). In the July 2012 letter St. Jude Medical informed customers that they would be implementing design improvements to the charger to address possible increased energy dissipation when the charger is misaligned with the IPG or the IPG is implanted too near the surface of the

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter