Integra Mozaik Strip 10 cc Product Usage: The product is a bone graft substitute composed of collagen (Type 1) and ceramic (I3-Tricalcium phosphate) matrix in putty and strip forms. The product is intended for use as a bone void filler to fill voids or gaps of the skeletal system in the extremities, spine. and pelvis not intrinsic to the stability of the bony structure. Integra Mozaik is also indicated for use in the treatment of surgically treated osseous defects or osseous defects created from traumatic injury to the bone. Following placement in the bone void or a defect Integra Mozaik is reabsorbed and replaced with bone during the healing process.
Publicado el 20 de junio de 2012 · Inicio del retiro: 11 de octubre de 2011
Resumen
FDA registra el retiro Z-1795-2012 de dispositivos médicos de Clase II, publicado el 20 de junio de 2012 e iniciado el 11 de octubre de 2011. Empresa responsable: Integra LifeSciences Corp.; Estado: Finalizado; Cantidad reportada: 22 units; Lote: Strip Part Number 7701513001 510 K- K063124 Device Listing Number - D001098 Integra MOZAIKTM Strip IOcc Reference number - CCM I021OITL Batch Number - 105BR0215955 Integra MOZAIKTM Strip 10cc Reference Number - CCM I021OITL Batch Number - 105BR0216149. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The Instructions for Use {IFU} {Putty Part Number 7700713001 and Strip Part Number 770151300l for CE marked Mozaik Strip and Putty products contains an incorrect translation of the term "bone marrow aspirate". The term is incorrectly translated in Spanish and French,
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Integra LifeSciences Corp.
- Distribuidor
- Worldwide Distribution
- Lote
- Strip Part Number 7701513001 510 K- K063124 Device Listing Number - D001098 Integra MOZAIKTM Strip IOcc Reference number - CCM I021OITL Batch Number - 105BR0215955 Integra MOZAIKTM Strip 10cc Reference Number - CCM I021OITL Batch Number - 105BR0216149
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDABrand Name: MediHoney® Gel Product Name: MediHoney® Wound & Burn Dressing Gel Model/Catalog Number: 3181519 de febrero de 2025
- FDASURG PAT XRAY 1/2X1-1/2 Model/Catalog Number: 801404. The surgical patties and surgical strips are indicated6 de noviembre de 2024
- FDASURG STRP 1-1/2X6 Model/Catalog Number: 801454. The surgical patties and surgical strips are indicated for the6 de noviembre de 2024
- FDASURG PAT XRAY 3/4X3/4 Model/Catalog Number: 801401. The surgical patties and surgical strips are indicated for6 de noviembre de 2024
- FDAPlasmaJet Open Surgery Handpiece (Part of the PlasmaJet Neutral Plasma Surgery System) The PlasmaJet system is20 de junio de 2012
- FDAIntegra Mozaik Putty 2.5cc, 5 cc, 10 cc Product Usage: The product is a bone graft substitute composed of20 de junio de 2012
Preguntas sobre este retiro
¿Integra Mozaik Strip 10 cc Product Usage: The product is a bone graft substitute composed of collagen (Type 1) and ceramic (I3-Tricalcium phosphate) matrix in putty and strip forms. The product is intended for use as a bone void filler to fill voids or gaps of the skeletal system in the extremities, spine. and pelvis not intrinsic to the stability of the bony structure. Integra Mozaik is also indicated for use in the treatment of surgically treated osseous defects or osseous defects created from traumatic injury to the bone. Following placement in the bone void or a defect Integra Mozaik is reabsorbed and replaced with bone during the healing process. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1795-2012 el 20 de junio de 2012. Peligro reportado: The Instructions for Use {IFU} {Putty Part Number 7700713001 and Strip Part Number 770151300l for CE marked Mozaik Strip and Putty products contains an incorrect translation of the term "bone marrow aspirate". The term is incorrectly translated in Spanish and French,
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter