Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1810-2012

T4-Hytrel Zipper Toga Product Usage - The togas are intended to provide a barrier between the operating room and the members of the surgical team in order to help protect against contamination and/or exposure of infectious body fluids and harmful microorganisms.

Publicado el 20 de junio de 2012 · Inicio del retiro: 1 de junio de 2012

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The clear tape, that is intended to aid in the prevention of patients fluids from contacting the surgeon/surgical staff through the sleeve seam of the toga, was not applied during the manufacturing process. Stryker has received 5 complaints associated with the defective seams which allowed bleed through of bodily fluids.

Qué hacer

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Detalles del retiro

Fabricante
Stryker Instruments Div. of Stryker Corporation
Distribuidor
Worldwide Distribution - US (nationwide) including the states of: AK, AL, AZ, CA, CO, DE, FL, GA, IA, ID, IL, IN, KY, LA, MA, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, WV and the country of SWITZERLAND
Lote
Part 0400-821-000, 0400-841-000,and 0400-851-000 all lots
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States