Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1817-2014

HydroFinity Hydrophilic guidewire; Nitinol Guidewire with Hydrophilic Coating, Sterile. 0.035 OD, Standard Shaft, Straight Tip, 260 CM Length Manufactured for Covidien, Plymouth, MN; Manufactured by: NDC Inc., Fremont, CA. Manufactured in Costa Rica The HydroFinity Hydrophilic guidewire facilitates the introduction and placement of catheters and interventional devices to the desired anatomical location during diagnostic or interventional procedures.

Publicado el 2 de julio de 2014 · Inicio del retiro: 28 de mayo de 2014

Resumen

FDA registra el retiro Z-1817-2014 de dispositivos médicos de Clase I, publicado el 2 de julio de 2014 e iniciado el 28 de mayo de 2014. Empresa responsable: Nitinol Devices and Components, Inc.; Estado: Finalizado; Cantidad reportada: 8485 in US, 9595 - ROW - total, all models; Lote: Model HPRS35260- all lots. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Reports of outer polymer jacket to the core wire being damaged during use on the HydroFinity Guidewire.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Nitinol Devices and Components, Inc.
Distribuidor
Worldwide Distribution - USA (nationwide) and Internationally to Italy, France, Norway, Sweden, Germany, Belgium, United Kingdom and Canada.
Lote
Model HPRS35260- all lots.
Clasificación de riesgo
Class I
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿HydroFinity Hydrophilic guidewire; Nitinol Guidewire with Hydrophilic Coating, Sterile. 0.035 OD, Standard Shaft, Straight Tip, 260 CM Length Manufactured for Covidien, Plymouth, MN; Manufactured by: NDC Inc., Fremont, CA. Manufactured in Costa Rica The HydroFinity Hydrophilic guidewire facilitates the introduction and placement of catheters and interventional devices to the desired anatomical location during diagnostic or interventional procedures. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1817-2014 el 2 de julio de 2014. Peligro reportado: Reports of outer polymer jacket to the core wire being damaged during use on the HydroFinity Guidewire.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter