da Vinci Xi EndoWrist Suction Irrigator, 8 mm instrument; PN 480299-03; and SOFTWARE, EMBEDDED RLS, IS4000, A70_P6_B440; PN 610092-440. General and Plastic Surgery: The EndoWrist® Suction Irrigator is designed to be used in conjunction with an Intuitive Surgical da Vinci Surgical System and compatible suction and irrigation sources and tubing sets for delivering fluid to the surgical site and for evacuation and aspiration of fluids. The instrument may also be used for retraction and blunt dissection of tissue. The instrument tip is blunt and intended to contact tissue.
Publicado el 19 de abril de 2017 · Inicio del retiro: 31 de marzo de 2017
Resumen
FDA registra el retiro Z-1819-2017 de dispositivos médicos de Clase II, publicado el 19 de abril de 2017 e iniciado el 31 de marzo de 2017. Empresa responsable: Intuitive Surgical, Inc.; Estado: Finalizado; Cantidad reportada: 15 x 6 packs (90 units); Lote: All. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Intuitive Surgical has become aware that in specific scenarios with system software P6 and the da Vinci Xi Suction Irrigator (PN 480299-03), users can experience unexpected motion of a system arm.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Intuitive Surgical, Inc.
- Distribuidor
- US Only - one location each in AL, CO, KS, NV, NY, and 2 in TX
- Lote
- All
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAda Vinci SP Instrument Arm Drape, SP, 1-PACK, Part Number: 430013-13, 430013-1522 de julio de 2026
- FDADa Vinci ASSY, PSS, SP1098, Part Number: 3806011 de julio de 2026
- FDABrand Name: Da Vinci 5 Product Name: ASSY,DV5 TOWER,IS5000 Model/Catalog Number: 380746 Software Version: N/A10 de junio de 2026
- FDAIntuitive 8 mm SureForm 30 Gray Reloads Reference Numbers: 48230M-05 and 48230M-0613 de mayo de 2026
- FDACLENZIderm M.D. Pore Therapy Salicylic Acid 2% Acne Treatment 5 fl. oz. (148 mL), Valeant Pharmaceuticals26 de abril de 2017
- FDANordson Micromedics FibriJet Endoscopic Catheter Introducer, Catalog Number SA-3445. Sterilized using Ethylene19 de abril de 2017
Preguntas sobre este retiro
¿da Vinci Xi EndoWrist Suction Irrigator, 8 mm instrument; PN 480299-03; and SOFTWARE, EMBEDDED RLS, IS4000, A70_P6_B440; PN 610092-440. General and Plastic Surgery: The EndoWrist® Suction Irrigator is designed to be used in conjunction with an Intuitive Surgical da Vinci Surgical System and compatible suction and irrigation sources and tubing sets for delivering fluid to the surgical site and for evacuation and aspiration of fluids. The instrument may also be used for retraction and blunt dissection of tissue. The instrument tip is blunt and intended to contact tissue. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1819-2017 el 19 de abril de 2017. Peligro reportado: Intuitive Surgical has become aware that in specific scenarios with system software P6 and the da Vinci Xi Suction Irrigator (PN 480299-03), users can experience unexpected motion of a system arm.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter