Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-1843-2012

Olympic Cool-Cap System is intended to provide treatment for neonatal hypoxic-ischemic encephalopathy (HIE). The Cool-Cap is indicated for use in full term infants (more than 36 week gestational age) with clinical evidence of moderate to severe HIE. The device provides selective head cooling with mild systemic hypothermia to prevent or reduce the severity of neurologic injury associated with HIE. It is a selective head cooling system for treatment of neonatal hypoxic-ischemic encephalopathy (HIE) in infants. A cooling water cap is placed on the infant's head and the body is warmed using radiant warmers. The Cool-Cap provides 72 hours of monitored cooling. The device is labeled in part: "***OLYMPIC COOL-CAP Cooling Unit***Model No.***Made in USA***Software in this product is Natus Medical Incorporated***"

Publicado el 27 de junio de 2012 · Inicio del retiro: 21 de mayo de 2012

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Olympic Cool-Cap Cooling Module's power supply failed during treatment. The power supply failure can happen without prior indication and results in the unavailability of the system to begin or continue cooling treatment.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Natus Medical Incorporated
Distribuidor
Worldwide Distribution -- USA (nationwide) including the states of : AK; AR; AZ; CA; CO; DE; FL; GA; HI; ID; IL; IN; KY; MA; MI; MN; MS; ND; NE; NJ; NY; OH; PA; SC; SD; TN; TX; and WA. and the countries of : Argentina; Australia; Austria; Canada; Egypt; Germany; Hong Kong; Indonesia; Iran; Italy; Kuwait; Poland; Russia; Saudi Arabia; Singapore; Slovenia; South Africa; South Korea; Spain; Taiwan; Thailand; Turkey; United Arab Emirates; and United Kingdom.
Lote
Cooling Serial Numbers: 1008; 1009; 1011; 1012; 1014; 1016; 1017; 1018; 1020; 1021; 1022; 1023; 1024; 1025; 1027; 1028; 1030; 1031; 1034; 1035; 1036; 1038; 1044; 1045; 1048; 1050; 1051; 1052; 1053; 1054; 1055; 1056; 1059; 1060; 1062; 1063; 1064; 1065; 1066; 1070; 1071; 1072; 1073; 1074; 1075; 1076; 1077; 1080; 1081; 1082; 1083; 1085; 1086; 1087; 1088; 1090; 1091; 1092; 1093; 1094; 1095; 1096; 1097; 1101; 1116; 1117; 1118; 1119; 1120; 1122; 1124; 1125; 1126; 1127; 1128; 1129; 1130; 1131; 1132; 1136; 1137; 1138; 1139; 1141; 7002; 7003; 7004; 7005; 7007; 7008; 7009; 7010; 7011; 7013; 7014; 7016; 7017; 7018; 7019; 7020; 7021; 7022; 7023; 7024; 7025; 7026; 7027; 7028; 7030; 7031; 7032; 7033; 7034; 7036; 7037; 7038; 7039; 7040; 7041; 7042; 7043; 7044; 7045; 7046; 7047; 7048; 7049; 7051; 7052; 7053; 7054; 7055; 7250; 7251; 7252; 7253; 7255; 7256; 7257; 7258; 7259; 7260; 7261; 7262; 7263; 7264; 7265; 7266; 7267; 7268; 7269; 7270; 7271; 7272; 7273; 7274; 7275; 7276; 7277; 7278; 7279; 7280; 7281; 7282; 7283; 7284; 7285; 7286; 7288; 7289; 7290; 7291; 7292; 7293; 7294; 7363; 7364; 7365; 7366; 7372; and 7374.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States