AirLife Isothermal Breathing Circuit, Infant Respiratory Circuit, Heated; an Rx device for single patient use; CareFusion, 22745 Savi Ranch Parkway, Yorba Linda, CA 92887 USA, Made in Mexico Respiratory breathing circuits are used with a ventilator when mechanical ventilation is administered to a patient.
Publicado el 11 de julio de 2012 · Inicio del retiro: 29 de mayo de 2012
Resumen
FDA registra el retiro Z-1899-2012 de dispositivos médicos de Clase I, publicado el 11 de julio de 2012 e iniciado el 29 de mayo de 2012. Empresa responsable: Carefusion 2200 Inc; Estado: Finalizado; Cantidad reportada: N/A; Lote: product codes 10127-4H1, 10331N-4S2, 10351-4H2, 10387-4S2, 10555-4S2, 10706-4S2, 10790-4S2, 10814-4S2, 10849-4S2, 1175-4H2, 1208-4H1, 1553-4H1, 1815-409, 1998-4H1, 2119-4S2G, 2120-4S2G, 4319-4H2, 4326-4H2, 4652-5014677-4H2, 4875-4H2, 5665-4H2, 5682-409, 5732-4H1, 5868-4H2, 6002-4H2, 6040-501, 6074-4H1, 6198-4H2, 6274-H12, 6301-H12, 6313-501, 6459-4H2; 6485-4H2, 6588-4H2, 6942-4H2, 7100-4S2, 7147-4S2, 7156-4S2, 7184-4S2, 7206-4S2, 7213-4S2, 7354-4S2, 7395-4S2, 7407-4S2, 7441-4S2, 7483-4S2, 7526-4S2, 7575-4S2, 7582-4S2, 7588-4S2, 9053-4S2, 9054-4S2, 9055-4S2, 9056-4S2, 9058-4S2, RC41-12027, RC41-18039, RC41-18052, RC41-18060, RC51-12074, RC51-18004, RT4851-00, RT4851-12 and RT4851-18; includes all lot numbers of the affected product codes manufactured prior to February 3, 2012, lot #0000414257 and any lot numbers beginning with the letter Y. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Carefusion 2200 Inc
- Distribuidor
- Worldwide distribution: USA (nationwide) including Puerto Rico and internationally to: Canada, Egypt, France, Germany, Great Britain, Italy, Kuwait, Mexico, Portugal, Saudi Arabia, Spain, Taiwan and United Arab Emirates.
- Lote
- product codes 10127-4H1, 10331N-4S2, 10351-4H2, 10387-4S2, 10555-4S2, 10706-4S2, 10790-4S2, 10814-4S2, 10849-4S2, 1175-4H2, 1208-4H1, 1553-4H1, 1815-409, 1998-4H1, 2119-4S2G, 2120-4S2G, 4319-4H2, 4326-4H2, 4652-5014677-4H2, 4875-4H2, 5665-4H2, 5682-409, 5732-4H1, 5868-4H2, 6002-4H2, 6040-501, 6074-4H1, 6198-4H2, 6274-H12, 6301-H12, 6313-501, 6459-4H2; 6485-4H2, 6588-4H2, 6942-4H2, 7100-4S2, 7147-4S2, 7156-4S2, 7184-4S2, 7206-4S2, 7213-4S2, 7354-4S2, 7395-4S2, 7407-4S2, 7441-4S2, 7483-4S2, 7526-4S2, 7575-4S2, 7582-4S2, 7588-4S2, 9053-4S2, 9054-4S2, 9055-4S2, 9056-4S2, 9058-4S2, RC41-12027, RC41-18039, RC41-18052, RC41-18060, RC51-12074, RC51-18004, RT4851-00, RT4851-12 and RT4851-18; includes all lot numbers of the affected product codes manufactured prior to February 3, 2012, lot #0000414257 and any lot numbers beginning with the letter Y.
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASnowden-Pencer MIS DIAMOND-TOUCH CLAMP DEBAKEY 35 MM JAW 5MM, 32CM, RATCHETED, REF SP90-7044; endoscopic23 de agosto de 2023
- FDACarefusion V. Mueller" CUSHING BAYONET BIPOLAR FORCEPS, 2.0MM TIP, INSULATED OVERALL LENGTH 7-1/2" (190MM),14 de diciembre de 2022
- FDACarefusion V. Mueller"GERALD BIPOLAR FORCEPS INSULATED, 1.0MM TIP OVERALL LENGTH 7" (180MM), REF F-111214 de diciembre de 2022
- FDACarefusion V. Mueller" HARDY-STYLE BAYONET BIPOLAR FORCEPS, 0.5MM TIP, INSULATED OVERALL LENGTH 8-1/2" (215MM)14 de diciembre de 2022
- FDANorthern Catch Harvest of the Sea Fancy Whole Oysters and SeaNet Fancy Whole Oysters; Net Wt. 8 oz (227g)11 de julio de 2012
- FDAAirLife Isothermal Breathing Circuit, Infant Respiratory Circuit, Non-Heated; an Rx device for single patient11 de julio de 2012
Preguntas sobre este retiro
¿AirLife Isothermal Breathing Circuit, Infant Respiratory Circuit, Heated; an Rx device for single patient use; CareFusion, 22745 Savi Ranch Parkway, Yorba Linda, CA 92887 USA, Made in Mexico Respiratory breathing circuits are used with a ventilator when mechanical ventilation is administered to a patient. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1899-2012 el 11 de julio de 2012. Peligro reportado: The Y adapter within the breathing circuit may spontaneously crack, causing a leak in the closed ventilation system. This leak could lead to a decrease in the tidal volume delivered to the ventilated patient.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter