Nestle Nutrition, Compat, Enteral Feeding Tube, sterile, 8 French (REF 08310400), 10 French (REF 08310600), and 12 French (REF 08310500). Product Usage: The Compat Nasogastric Feeding Tube is indicated for the administration of nutrition, fluids and medications into the stomach r small bowel by the naso-enteric, route for patients who have an intact gastrointestinal tract but are physically unable to ingest enough nutrients orally to meet their current needs.
Publicado el 14 de agosto de 2013 · Inicio del retiro: 24 de julio de 2013
Resumen
FDA registra el retiro Z-1922-2013 de dispositivos médicos de Clase II, publicado el 14 de agosto de 2013 e iniciado el 24 de julio de 2013. Empresa responsable: Nestle HealthCare Nutrition; Estado: Finalizado; Cantidad reportada: 37,310; Lote: REF 08310400 - Compat® Nasogastric Feeding Tube 8 French (8Fr) in the following lots: 01413G 02412K 08211Z 08712W 14212U 14611P 21512Y 35411H REF 08310500- Compat® Nasogastric Feeding Tube 12 French (12Fr) in the following lots: 01211AV 01513N 04111D, 04111P, 04111V 05812F 05912F 06313N 21512G 23412BV 34712V 35411U REF 08310600 - Compat® Nasogastric Feeding Tube 10 French (10Fr) in the following lots: 08111M 08712D 10211E 13712W 21512F 05312N. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Nestle HealthCare Nutrition, Inc. is initiating a recall on some of Compat NG Tubes because they have been delivered in an insufficiently sealed pouch, thus compromising the sterility claimed on the label.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Nestle HealthCare Nutrition
- Distribuidor
- US Nationwide Distribution - including the states of: AK, AZ, CA, CO, CT, FL, GA, HI, IA, IL, IN, KS, KY, MA, MD, MI, MN, MO, NC, NE, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA.
- Lote
- REF 08310400 - Compat® Nasogastric Feeding Tube 8 French (8Fr) in the following lots: 01413G 02412K 08211Z 08712W 14212U 14611P 21512Y 35411H REF 08310500- Compat® Nasogastric Feeding Tube 12 French (12Fr) in the following lots: 01211AV 01513N 04111D, 04111P, 04111V 05812F 05912F 06313N 21512G 23412BV 34712V 35411U REF 08310600 - Compat® Nasogastric Feeding Tube 10 French (10Fr) in the following lots: 08111M 08712D 10211E 13712W 21512F 05312N
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAbioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit V2S14 de agosto de 2013
- FDABD Vacutainer Multiple Sample and Direct Draw Luer Adapter BD Franklin Lakes, NJ 07417 USA A sterile14 de agosto de 2013
- FDAbioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDT14 de agosto de 2013
- FDADimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (DC107) Product Usage: The TACR14 de agosto de 2013
- FDAbioMerieux Vitek 2 Systems Software version 6.01 software, used with the VITEK 2 automated system. Kit, UPDT14 de agosto de 2013
- FDADimension(R) Mass Creatine Kinase MB Isozyme Calibrator (RC420) The Mass Creatine Kinase MB isozyme Calibrator14 de agosto de 2013
Preguntas sobre este retiro
¿Nestle Nutrition, Compat, Enteral Feeding Tube, sterile, 8 French (REF 08310400), 10 French (REF 08310600), and 12 French (REF 08310500). Product Usage: The Compat Nasogastric Feeding Tube is indicated for the administration of nutrition, fluids and medications into the stomach r small bowel by the naso-enteric, route for patients who have an intact gastrointestinal tract but are physically unable to ingest enough nutrients orally to meet their current needs. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1922-2013 el 14 de agosto de 2013. Peligro reportado: Nestle HealthCare Nutrition, Inc. is initiating a recall on some of Compat NG Tubes because they have been delivered in an insufficiently sealed pouch, thus compromising the sterility claimed on the label.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter