Dimension Tacrolimus Flex reagent cartridge (DF107) and Dimension TACR CAL (DC107) Product Usage: The TACR method is an in vitro diagnostic test intended to quantitatively measure Tacrolimus in human whole blood on the Dimension clinical chemistry system. Measurements of Tacrolimus are used as an aid in the management of tacrolimus therapy in kidney and liver transplant patients.

Publicado el 14 de agosto de 2013 · Inicio del retiro: 1 de julio de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Siemens initiated a recall due to confirmed complaints of low patient sample recovery with Dimension TACR Flex Reagent cartridge (DF107) lot BB4087 and its linked calibrator (DC 107) lot 3BD029. Falsely low Tacrolimus results across the assay range may lead to alterations in dosing.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Siemens Healthcare Diagnostics, Inc.
Distribuidor
USA Nationwide Distribution including the states of: TX, NC, GA, TN, UT, NJ, MI, VA, AR, LA, MO, FL, AZ, CA, DC, PA, NH, WI, SD, SC, NM, MD, and IN.
Lote
TACR Flex Reagent cartridge, material #10444938, (DF107) lot BB4087 with expiration date 3/28/14 and its linked calibrator, material #10445012, (DC 107) lot 3BD029 with expiration date 03/01/14.
Clasificación de riesgo
Class II