AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, including single use indications, is maintained for each device, which may be used independently and at different times. These specific kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that were recalled as part of RES 98317 with (510(k): K140100).
Publicado el 6 de mayo de 2026 · Inicio del retiro: 27 de marzo de 2026
Resumen
FDA registra el retiro Z-1947-2026 de dispositivos médicos de Clase I, publicado el 6 de mayo de 2026 e iniciado el 27 de marzo de 2026. Empresa responsable: North American Rescue LLC.; Estado: En curso; Cantidad reportada: 21; Lote: Kit REF/UDI-DI/Lot: 85-4661/00842209132089/85-4661122024; 85-4742/00842209133536/85-4742123024. Meter: UDI-DI: 20612479197217, all lots. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.
Qué hacer
Voluntary: Firm initiated
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- North American Rescue LLC.
- Distribuidor
- US: CA, WA
- Lote
- Kit REF/UDI-DI/Lot: 85-4661/00842209132089/85-4661122024; 85-4742/00842209133536/85-4742123024. Meter: UDI-DI: 20612479197217, all lots.
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDABrand Name: Disposable Mixing Bowls with Spatula Product Name: Mixing Bowl and Spatula Model/Catalog Number:6 de mayo de 2026
- FDAMedtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment6 de mayo de 2026
- FDAJUNCTIONAL EMERGENCY TREATMENT TOOL (JETT), REF: 30-0088, a component inside kits: KIT, TCCC Training-8 de octubre de 2025
- FDARange Trauma Kit ORG - Product Code 80-0213, 80-0299, 80-0298, North American Rescue14 de diciembre de 2016
- FDAUSCG Boat Response Kit - Product Code 80-0353, North American Rescue14 de diciembre de 2016
- FDARange Trauma Kit Hardcase - Product Code 85-0889, North American Rescue14 de diciembre de 2016
Preguntas sobre este retiro
¿AIDBAGs are first aid kits of convenience composed of individually labeled devices designed as specialty first aid kits designed for the appropriate level provider to evaluate and treat patients at the point-of-injury / point-of-illness. The original manufacturer labeling, including single use indications, is maintained for each device, which may be used independently and at different times. These specific kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that were recalled as part of RES 98317 with (510(k): K140100). fue retirado del mercado?
Sí. FDA publicó el retiro Z-1947-2026 el 6 de mayo de 2026. Peligro reportado: Kits contain the McKesson TRUE Metrix PRO Professional Monitoring Blood Glucose System that has been recalled for an issue with the software design of the E-5 Error Code where the meter displays an E-5 error code for a very high blood glucose event (> 600 mg/dL) or when there is a test strip error, which can lead to operator confusion when evaluating patients and a delay in appropriate treatment. This is a downstream recall of Trividia Health Class I, Recall Event ID 98317.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated