FDAFDA · Alimentos y medicamentos·N.º oficial Z-1962-2013

Remel X/pect Flu A&B Control Swabs, For In Vitro Diagnostic Use, Ref R246003, each package contains 20 Flu A Control Swabs and 20 Flu B Control Swabs, each swab is individually packaged in a foil pouch. The firm name on the label is Remel Europe, Ltd., Dartford, Kent, United Kingdom. Product is a test kit containing characterized positive and negative controls intended for use in qualitative procedures as external quality control material for the Remel Xpect Flu A&B test kit.

Publicado el 21 de agosto de 2013 · Inicio del retiro: 19 de febrero de 2013

Resumen

FDA registra el retiro Z-1962-2013 de dispositivos médicos de Clase III, publicado el 21 de agosto de 2013 e iniciado el 19 de febrero de 2013. Empresa responsable: Remel Inc; Estado: Finalizado; Cantidad reportada: 96 kits of 20 swab sets; Lote: Lot 260745, exp. 5/31/2014. La Clase III indica que no es probable que el producto cause consecuencias adversas para la salud.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Flu A+ Control Swabs incorrectly give Flu A- results when used for quality control testing with the Xpect Flu A&B test kit.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Remel Inc
Distribuidor
Nationwide distribution: AL, AR, CA, CO, CT, FL, GA, IA, IL, KS, MA, MD, MI, MO, NM, NV, NY, OH, PA, TX, WA, and WI.
Lote
Lot 260745, exp. 5/31/2014
Clasificación de riesgo
Class III
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Remel X/pect Flu A&B Control Swabs, For In Vitro Diagnostic Use, Ref R246003, each package contains 20 Flu A Control Swabs and 20 Flu B Control Swabs, each swab is individually packaged in a foil pouch. The firm name on the label is Remel Europe, Ltd., Dartford, Kent, United Kingdom. Product is a test kit containing characterized positive and negative controls intended for use in qualitative procedures as external quality control material for the Remel Xpect Flu A&B test kit. fue retirado del mercado?

Sí. FDA publicó el retiro Z-1962-2013 el 21 de agosto de 2013. Peligro reportado: Flu A+ Control Swabs incorrectly give Flu A- results when used for quality control testing with the Xpect Flu A&B test kit.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter