enFlow IV Fluid Warmer (510k name: EFLOW IV FLUID WARMER, MODEL 100, EFLOW DISPOSABLE CARTRIDGE, MODEL 200 AND EFLOW AC POWER SUPPLY, MODEL 120). The Enginivity enFlow TM IV Fluid Warmer is intended for warming blood, blood products and intravenous solutions prior to administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia.
Publicado el 18 de julio de 2012 · Inicio del retiro: 20 de febrero de 2012
Resumen
FDA registra el retiro Z-1972-2012 de dispositivos médicos de Clase II, publicado el 18 de julio de 2012 e iniciado el 20 de febrero de 2012. Empresa responsable: Vital Signs Devices, a GE Healthcare Company; Estado: Finalizado; Cantidad reportada: 11803 cases; Lote: a) 510(k): K060537 b) Medical device listing number: D071942 All enFlow Warmer Straps in the field present a potential for tissue injury. Sold separately as Product #: 980304 and 980304EU Included as an accessory in following enFlow products: Military Domestic 91000153 980100 91000154 980100EU 91000156 980105VS 91000103. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The EnFlow Warmer Strap does not meet the requirements of the biocompatibility standard (ISO 10993) for products that may have contact with skin for less than 24 hours.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Vital Signs Devices, a GE Healthcare Company
- Distribuidor
- Nationwide Distribution.
- Lote
- a) 510(k): K060537 b) Medical device listing number: D071942 All enFlow Warmer Straps in the field present a potential for tissue injury. Sold separately as Product #: 980304 and 980304EU Included as an accessory in following enFlow products: Military Domestic 91000153 980100 91000154 980100EU 91000156 980105VS 91000103
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDADisposable Temperature Probes, (sterile), Sensor diameter: 3mm/0.12 in; 25 units per box. Vital Signs Inc. a21 de agosto de 2013
- FDACorometrics Qwik Connect Plus Spiral Electrode- Model #7000AA0 A fetal scalp circular (spiral) electrode and28 de noviembre de 2012
- FDASystem 1 Base 220/240V. The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the18 de julio de 2012
- FDAiStat1 Analyzer Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests for sodium,18 de julio de 2012
- FDASystem 1 Base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use for up to 6 hours on the18 de julio de 2012
- FDAi-Stat Portable Clinical Analyzer; Abbott Point of Care, Princeton, NJ 08540 Model 200. The 37 degree tests18 de julio de 2012
Preguntas sobre este retiro
¿enFlow IV Fluid Warmer (510k name: EFLOW IV FLUID WARMER, MODEL 100, EFLOW DISPOSABLE CARTRIDGE, MODEL 200 AND EFLOW AC POWER SUPPLY, MODEL 120). The Enginivity enFlow TM IV Fluid Warmer is intended for warming blood, blood products and intravenous solutions prior to administration. It is intended to be used by healthcare professionals in hospital, clinical and field environments to help prevent hypothermia. fue retirado del mercado?
Sí. FDA publicó el retiro Z-1972-2012 el 18 de julio de 2012. Peligro reportado: The EnFlow Warmer Strap does not meet the requirements of the biocompatibility standard (ISO 10993) for products that may have contact with skin for less than 24 hours.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter