Medtronic MiniMed Paradigm Reservoirs, Model No: MMT-326A (1.8mL) and MMT-332A (3.0mL) The model MMT-326A and MMT-332A are syringe type insulin reservoirs intended for use with Medtronic Paradigm series insulin infusion pumps. The two reservoirs are identical except for barrel length and as a result, maximum insulin capacity. The MMT-326A holds a maximum of 1.8ml of insulin whereas the longer MMT-332A reservoir can hold up to 3.0ml of insulin. Both reservoirs are supplied sterile and are labeled for single use. Typically, the reservoir is replaced every tow or three days at the time of infusion set replacement. The Paradigm reservoirs have a shelf life of three years post sterilization. Product Quantity Distributed (Int) 669,672 units
Publicado el 28 de agosto de 2013 · Inicio del retiro: 3 de julio de 2013
Resumen
FDA registra el retiro Z-1986-2013 de dispositivos médicos de Clase II, publicado el 28 de agosto de 2013 e iniciado el 3 de julio de 2013. Empresa responsable: Medtronic MiniMed; Estado: Finalizado; Cantidad reportada: 669,672 units; Lote: MMT-326A Lot Numbers: H8437486 H8441420 H8442973 H8451531 H8486688 H8489386 H8485398 H8500423 H8500472 H8503728 H8512826 H8503372 H8510440 H8539013 H8627745 MMT-332A Lot Numbers: H8416432 H8420977 H8422490 H8424676 H8452933 H8455959 H8457716 H8459557 H8461538 H8463297 H8464121 H8467888 H8469703 H8471745 H8473271 H8476270 H8478398 H8473106 H8492449 H8491921 H8494645 H8496561 H8509305 H8512566 H8515317 H8517079 H8521052 H8541843 H8584244 H8603292 H8604958 H8635301. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Medtronic is recalling certain lots of Medtronic MiniMed Paradigm Reservoirs MMT-326A (1.8mL) and MMT-332A (3.0mL) used with Medtronic Paradigm insulin pumps because they may have increased risk for leaking. A leak in the reservoir may result in delivery of less insulin than intended. In addition, if there is a leaky reservoir and an insulin blockage occurs in the infusion set, the pump may not
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Medtronic MiniMed
- Distribuidor
- Worldwide Distribution - USA (nationwide) and the countries of Australia, Bermuda, Brazil, Canada, Chile, Columbia, Ecuador, El Salvador, Germany, Japan, Mexico, Netherland, Paraguay, Puerto Rico, Singapore, Uruguay and Venezuela.
- Lote
- MMT-326A Lot Numbers: H8437486 H8441420 H8442973 H8451531 H8486688 H8489386 H8485398 H8500423 H8500472 H8503728 H8512826 H8503372 H8510440 H8539013 H8627745 MMT-332A Lot Numbers: H8416432 H8420977 H8422490 H8424676 H8452933 H8455959 H8457716 H8459557 H8461538 H8463297 H8464121 H8467888 H8469703 H8471745 H8473271 H8476270 H8478398 H8473106 H8492449 H8491921 H8494645 H8496561 H8509305 H8512566 H8515317 H8517079 H8521052 H8541843 H8584244 H8603292 H8604958 H8635301
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAMedtronic MiniMed Paradigm Insulin pump, model number MMT-523K1 de abril de 2020
- FDAMiniMed Insulin Pump, model # MMT-5081 de abril de 2020
- FDAMedtronic MiniMed Paradigm Insulin pump, model number MMT-5221 de abril de 2020
- FDAMedtronic MiniMed Paradigm Insulin pump, model number MMT-7231 de abril de 2020
- FDAEdwards Lifesciences Suction Wand Model S099B, packaged in Sorin Group Heart/Lung Perfusion Packs.28 de agosto de 2013
- FDAPleuraGuide Disposable Chest Tube Kit; Product codes: 17100, 17150, and 17200. Indicated to assist with the28 de agosto de 2013
Preguntas sobre este retiro
¿Medtronic MiniMed Paradigm Reservoirs, Model No: MMT-326A (1.8mL) and MMT-332A (3.0mL) The model MMT-326A and MMT-332A are syringe type insulin reservoirs intended for use with Medtronic Paradigm series insulin infusion pumps. The two reservoirs are identical except for barrel length and as a result, maximum insulin capacity. The MMT-326A holds a maximum of 1.8ml of insulin whereas the longer MMT-332A reservoir can hold up to 3.0ml of insulin. Both reservoirs are supplied sterile and are labeled for single use. Typically, the reservoir is replaced every tow or three days at the time of infusion set replacement. The Paradigm reservoirs have a shelf life of three years post sterilization. Product Quantity Distributed (Int) 669,672 units fue retirado del mercado?
Sí. FDA publicó el retiro Z-1986-2013 el 28 de agosto de 2013. Peligro reportado: Medtronic is recalling certain lots of Medtronic MiniMed Paradigm Reservoirs MMT-326A (1.8mL) and MMT-332A (3.0mL) used with Medtronic Paradigm insulin pumps because they may have increased risk for leaking. A leak in the reservoir may result in delivery of less insulin than intended. In addition, if there is a leaky reservoir and an insulin blockage occurs in the infusion set, the pump may not
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter