FDAFDA · Alimentos y medicamentos·N.º oficial Z-1988-2020
Dover 100% Silicone Premium Urine Meter Foley Tray, 400mL, 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement, Sterile, Product Code: P4P14TSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.
Publicado el 20 de mayo de 2020 · Inicio del retiro: 30 de abril de 2020
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Temperature sensing catheters failed to meet the established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient body temperatures could lead to an incorrect diagnosis and insufficient treatment
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Covidien Llc
- Distribuidor
- US Nationwide distribution.
- Lote
- Lot Numbers: 1909308464, 1933000164
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?