FDAFDA · Alimentos y medicamentos·N.º oficial Z-2006-2012
CareFusion EnVe Ventilator Designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs). It is suitable for service in hospital and transport environments as a source of continuous or intermittent positive pressure ventilatory support, delivered invasively or non-invasively. It is not intended for homecare use.
Publicado el 25 de julio de 2012 · Inicio del retiro: 19 de junio de 2012
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
CareFusion identified that the EnVe Ventilator may not hold the set Positive End Expiratory Pressure (PEEP) value either intermittently or continuously or it may alarm for a Patient Circuit Fault. The unit will activate either a Low PEEP or Patient Circuit Fault Alarm both audibly and Visually to alert the healthcare professional. Ventilation delivery to the patient may be compromised.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- CareFusion 203, Inc.
- Distribuidor
- Nationwide Distribution including AZ, CA, CO, FL, IL, KY, ME, MI, MO, MN, MT, ND, NJ, NM, NV, NY, OK, PA, SC, SD, TX , UT, WI, and WV.
- Lote
- serial numbers: 26562, 43634, 43642, 43643, 44536, 44537, 44540, 44541, 44542, 44543, 44544, 44545, 44546, 44552, 44553, 44554, 44555, 44557, 44558, 44559, 44560, 44561, 44562, 44563, 44564, 44565, 45318, 45320, 45321, 45322, 45323, 45324, 45325, 45326, 45327, 45328, 45329, 45330, 45333, 45334, 45335, 45336, 45337, 45338, 45339, 45340, 45341, 45342, 45343, 45344, 45345, 45346, 45735, 45736, 45737, 45738, 45739, 45740, 45741, 45742, 45743, 45744, 45745, 45746, 45748, 45750, 45751, 45752, 45753, 45754, 45902, 45903, 45904, 45905, 45906, 45907, 45908, 45909, 45910, 45911, 46038, 46039, 46040, 46041, 46042, 46043, 46044, 46045, 46046, 46047, 46048, 46049, 46050, 46051, 46052, 46632, 46634, 46635, 46636, 46637, 46638, 46639, 46640, 46641, 46642, 46643, 46644, 46645, 47069, 47070, 47071, 47073, 47074, 47075, 47076, 47077, 47078, 47079, 47080, 47081, 47082, 47083, 47210, 47211, 47212, 47213, 47214, 47215, 47216, 47217, 47219, 47221, 47222, 47223, 47225, 47226, 47227, 47229, 47230, 47231, 47232, 47233, 47234, 47236, 47237, 47238, 47239, 47538, 47539, 47540, 47541, 47542, 47543, 47544, 47545, 47546, 47547, 47548, 47549, 47550, 47551, 47552, 47553, 47554, 47555, 47556, 47557, 47558, 47559, 47560, 47561, 47562, 47563, 47564, 47565, 47566, 47567, 48759, 48760, 48761, 48762, 48763, 48764, 48765, 48766, 48767, 48768, 48769, 48770, 48771, 48772, 48773, 48774, 48775, 48776, 48777, 48778, 48779, 48780, 48781, 48782, 48783, 48784, 48785, 48786, 48787, 48788, 48868, 48869, 48870, 48871, 48872, 48873, 48874, 48875, 48876, 48878, 48881, 48883, 48884, 48885, 48886, 48887, 48889, 48890, 48891, 48892, 48893, 48894, 48895, 48897, 50390, 50391, 50392, 50393, 50394, 50395, 50397, 50409, 50411, 50412, 50416, 50417, 50418, 50419, 50648, 50650, 50651, 50652, 50654, 50656, 50657, 50658, 50659, 50660, 50661, 50662, 50663, 50664, 50665, 50666, 50667, 50668, 50672, 51569, 51574
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States