Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2022-2014

STOPCOCK EXT. SET W/ 3-WAY STOPCOCK WITH INJ. SITE AND SPINLOCK CONNECTOR, 22 inches and 37 inches

Publicado el 16 de julio de 2014 · Inicio del retiro: 29 de mayo de 2014

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Complaints were received that reported the stopcock of some Stopcock Extension Sets are assembled in the reverse position, resulting in the stopcock being closed when set in the open position and vise versa.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
B. Braun Medical, Inc.
Distribuidor
USA (nationwide) including the states of PA, IL, NC, FL, IL, CA, CO, NE, LA, NC, LA, VA, TN, NE, TX, and KS.
Lote
Product codes: 473547 and 473549 lot #s and expiration dates: 61356774, 1/31/2019; 61362114, 2/28/2019; and 61367665, 3/31/2019.
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States