FDAFDA · Alimentos y medicamentos·N.º oficial Z-2022-2014
STOPCOCK EXT. SET W/ 3-WAY STOPCOCK WITH INJ. SITE AND SPINLOCK CONNECTOR, 22 inches and 37 inches
Publicado el 16 de julio de 2014 · Inicio del retiro: 29 de mayo de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Complaints were received that reported the stopcock of some Stopcock Extension Sets are assembled in the reverse position, resulting in the stopcock being closed when set in the open position and vise versa.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- B. Braun Medical, Inc.
- Distribuidor
- USA (nationwide) including the states of PA, IL, NC, FL, IL, CA, CO, NE, LA, NC, LA, VA, TN, NE, TX, and KS.
- Lote
- Product codes: 473547 and 473549 lot #s and expiration dates: 61356774, 1/31/2019; 61362114, 2/28/2019; and 61367665, 3/31/2019.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?
Recalert vigila todo tu catálogo contra retiros como este, automáticamente.