SHAPE ARM, SINGLE, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as a component of kit part number 0682400253 and the Arm part number in the kit is P17085. . The kit, when assembled, supports a flat-panel video monitor that is attached to one of two types of Stryker Communications boom systems, EDS or FLEXiS.
Publicado el 23 de julio de 2014 · Inicio del retiro: 17 de marzo de 2014
Resumen
FDA registra el retiro Z-2052-2014 de dispositivos médicos de Clase II, publicado el 23 de julio de 2014 e iniciado el 17 de marzo de 2014. Empresa responsable: Stryker Communications; Estado: Finalizado; Cantidad reportada: 953 STRYKER SHAPE ARMS of all types; Lote: Serial numbers 111018DE001 through 140210DE020 Serial and part numbers are located on underside of the cover of the portion of the Arm that is attached to the boom's service head. Screws must be removed and the cover detached to view these numbers. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
During the use of the SHAPE Arm there is the potential for the monitor mount to come off of the SHAPE Arm if the lock washer and screw are not present.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Stryker Communications
- Distribuidor
- Worldwide Distribution. US nationwide including Puerto Rico, China, Philippines, Mexico, Korea, Japan, Colombia, Chile, Canada, Brazil, and Australia
- Lote
- Serial numbers 111018DE001 through 140210DE020 Serial and part numbers are located on underside of the cover of the portion of the Arm that is attached to the boom's service head. Screws must be removed and the cover detached to view these numbers.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAChromophare Surgical Light System, REF: CH00000001; SLX to Oculan NFC Upgrade Kit, REF: P600341 de octubre de 2025
- FDAChromophare Surgical Light System , REF CH0000000117 de abril de 2024
- FDABerchtold Chromophare Dual Monitor Carrier Arm or Dual Flat Panel, a component of Berchtold Chromophare11 de septiembre de 2019
- FDAEDS Light Suspension, Central Axis, a component of the Stryker Visum LED Surgical Lighting System, with the29 de mayo de 2019
- FDAShape HF Cardiopulmonary System. Made up of Shape HF System Analyzer (PN 0001-9001) and Disposable Patient23 de julio de 2014
- FDASHAPE ARM, DUAL SH2, an articulating, arm bracket that is attached to the service head of boom systems which23 de julio de 2014
Preguntas sobre este retiro
¿SHAPE ARM, SINGLE, an articulating, arm bracket that is attached to the service head of boom systems which are used for supporting and positioning equipment in operating rooms and patient care areas. The product comes as a component of kit part number 0682400253 and the Arm part number in the kit is P17085. . The kit, when assembled, supports a flat-panel video monitor that is attached to one of two types of Stryker Communications boom systems, EDS or FLEXiS. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2052-2014 el 23 de julio de 2014. Peligro reportado: During the use of the SHAPE Arm there is the potential for the monitor mount to come off of the SHAPE Arm if the lock washer and screw are not present.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter