FDAFDA · Alimentos y medicamentos·N.º oficial Z-2062-2016

*** 1) Product labeled in part: 19683-28; LIFESHIELD; MICRODRIP SOLUSET WITH 1.2 MICRON FILTER 150ML BURETTE PLUMSET, CONVERTIBLE PIN, CLAVE SECONDARY PORT, OPTION-LOK, NON-DEHP; NOMINAL LENGTH: 118 IN; PRIMING VOLUME: 14 ML; *** 2) Product labeled in part: List No. 19685-28; LifeShield; LATEX-FREE; PLUMSET, PRIMARY CONVERTIBLE PP W/ SIGHT CHAMBER, THREE WAY STOPCOCK, CLAVE INJECTION PORTS, OPTION-LOK, NON-DEHP; NOMINAL LENGTH: 136 IN (345 CM); PRIMING VOLUME: 24 ML; *** 3) Product labeled in part: 19720-28; LIFESHIELD; PLUM SET, CONV P.P, DUAL CHANNEL 3 CLAVE PORTS,IN LINE CHECK VAL OPTION LOK, NON-DEHP.; NOMINAL LENGTH:145 IN (368 CM); PRIMING VOLUME:13 ML; *** 4) Product labeled in part: 19721-28; LIFESHIELD; SOLUSET 150ML BUR,CONV P.P DUAL CHANNEL 4 CLAVE PORTS IN LINE CHECK VAL OPTION LOK, NON-DEHP; NOMINAL LENGTH: 145 IN (368 CM); PRIMING VOLUME: 15 ML: *** 5) Product labeled in part: 19729-28; LIFESHIELD; CONV. DUAL CHANNEL PLUMSET WITH CLAVES, AND SECURE LOK, NON-DEHP; NOMINAL LENGTH: 96 IN (244 CM); PRIMING VOLUME: 18 ML; *** 6) Product labeled in part: 19732-28; LIFESHIELD; LIFESHIELD LATEX-FREE PLUM SET,MICRODRIP,PP CHECK VALVE AND CLAVE OPTION-LOK, NON-DEHP; NOMINAL LENGTH:104 IN (264 CM); PRIMING VOLUME: 13 ML; Product Usage: The CLAVE of the secondary port is used as an easy secondary access that allows concurrent and piggyback mode in infusion. It is normally used with a syringe or a secondary IV infusion set.

Publicado el 6 de julio de 2016 · Inicio del retiro: 9 de septiembre de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Intravenous fluid may leak at the CLAVE secondary port of Hospira PlumSet IV administration sets.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Hospira Inc.
Distribuidor
US (Nationwide) Distribution
Lote
X
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States