Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2134-2013

CONFIDENCE SPINAL CEMENT SYSTEM KIT, Product Code: 2839-13-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures.

Publicado el 11 de septiembre de 2013 · Inicio del retiro: 17 de abril de 2013

Resumen

FDA registra el retiro Z-2134-2013 de dispositivos médicos de Clase II, publicado el 11 de septiembre de 2013 e iniciado el 17 de abril de 2013. Empresa responsable: DePuy Spine, Inc.; Estado: Finalizado; Cantidad reportada: 3896 units (Overall quantity distributed appears greater than quantities manufactured due to the distribution pathways); Lote: HPBB0B, HPBBWM, HPCB75, HPCCBN, HPBB0R, HPBBWP, HPCB7D, HPDBBV, HPBB28, HPBBZ7, HPCB7F, HPDBCG, HPBB2D, HPBBZF, HPCB7T, HPDBD1, HPBB2F, HPBCKH, HPCB7V, HPDBD5, HPBB4R, HPBDL9, HPCB8G, HPDBF9, HPBBV1, HPCB67, HPCBLC, HPDBGH, HPBBWG, HPCB68, HPCBM1, HPDBGK, HPBBWK, HPCB74, HPCCBM, HPDBGL. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
DePuy Spine, Inc.
Distribuidor
Worldwide Distribution - USA including AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IN, KY, LA, MA, MD, MI, MN, MO, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WV, WY; Worldwide: Algeria, Australia, Austria, Belgium, Brazil, Canada, Colombia, Cyprus, Czech Republic, France, Germany, Hong Kong, India, Israel, Italy, Latvia, Luxembourg, Malaysia, Mexico, Netherlands, Norway, Poland, Portugal, Singapore, Slovenia, South Africa, Spain, Sweden, Switzerland, Turkey, UAE, and UK
Lote
HPBB0B, HPBBWM, HPCB75, HPCCBN, HPBB0R, HPBBWP, HPCB7D, HPDBBV, HPBB28, HPBBZ7, HPCB7F, HPDBCG, HPBB2D, HPBBZF, HPCB7T, HPDBD1, HPBB2F, HPBCKH, HPCB7V, HPDBD5, HPBB4R, HPBDL9, HPCB8G, HPDBF9, HPBBV1, HPCB67, HPCBLC, HPDBGH, HPBBWG, HPCB68, HPCBM1, HPDBGK, HPBBWK, HPCB74, HPCCBM, HPDBGL
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿CONFIDENCE SPINAL CEMENT SYSTEM KIT, Product Code: 2839-13-000 The CONFIDENCE Spinal Cement System is intended for percutaneous delivery of CONFIDENCE Spinal Cement, which is indicated for fixation of pathological fractures of the vertebral body during vertebroplasty or kyphoplasty procedures. fue retirado del mercado?

Sí. FDA publicó el retiro Z-2134-2013 el 11 de septiembre de 2013. Peligro reportado: During injection of cement, the water in the hydraulic pump leaks past the piston within the pump body resulting in the loss of pressure and inability to continue to inject cement.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter