Patient Data Manager 2.0 (Content manager 2.0, Patient Browser 4.0, DICOM Viewer 2.0) Brainlab Digital Lightbox. A system for the display of medical images. Model numbers 25100-05 and 25100-06. The software can transfer images to and from picture archiving and communication systems (PACS), file servers, or removable storage media.
Publicado el 6 de agosto de 2014 · Inicio del retiro: 27 de junio de 2014
Resumen
FDA registra el retiro Z-2134-2014 de dispositivos médicos de Clase II, publicado el 6 de agosto de 2014 e iniciado el 27 de junio de 2014. Empresa responsable: Brainlab AG; Estado: Finalizado; Cantidad reportada: 154 systems; Lote: Model/Catalog Numbers: 25100-05 PATIENT DATA MANAGER 2.0.11 25100-06 PATIENT DATA MANAGER 2.0.2. Software Versions: 2.0.0, 2.0.1, and 2.0.2. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
The following defects have been identified for data loaded with Patient Data Manager version 2.0 (including subversions 2.0.0, 2.0.1 and 2.0.2): 1) When loading non-square pixel image data with PDM 2.0, the data set is not correctly transferred and displayed in the corresponding Brainlab navigation or planning software. When a data set is loaded, the pixels of the data set are sorted in order to
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Brainlab AG
- Distribuidor
- Worldwide Distribution-USA (nationwide) including the states of AL, AK, AZ, CA, FL, GA, IA, IL, KS, MD, MO, MS, NC, NY, OH, PA, SD, TX, VA, VT, WA, and WI, and the countries of Australia, Austria, Belgium, Brazil, Chile, China, Ecuador, France, Germany, Japan, Netherlands, New Zealand, Norway, Qatar, Russia, Spain, Switzerland, Taiwan, United Arab Emirates, and United Kingdom.
- Lote
- Model/Catalog Numbers: 25100-05 PATIENT DATA MANAGER 2.0.11 25100-06 PATIENT DATA MANAGER 2.0.2. Software Versions: 2.0.0, 2.0.1, and 2.0.2
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- Germany
Retiros relacionados
- FDAUltrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x and14 de octubre de 2020
- FDASpine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 )24 de abril de 2019
- FDAExacTrac versions 6.5 through 6.5 intended to be used to place patients at an accurately defined point within21 de marzo de 2018
- FDAAIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO is intended to be used for X-ray14 de febrero de 2018
- FDACephalexin Capsules, USP, 250 mg, packaged in a) 100-count bottles (NDC 62756-293-88) and b) 500-count bottles13 de agosto de 2014
- FDADimension® Integrated Chemistry Systems intended to measure a variety of analytes in human body fluids.6 de agosto de 2014
Preguntas sobre este retiro
¿Patient Data Manager 2.0 (Content manager 2.0, Patient Browser 4.0, DICOM Viewer 2.0) Brainlab Digital Lightbox. A system for the display of medical images. Model numbers 25100-05 and 25100-06. The software can transfer images to and from picture archiving and communication systems (PACS), file servers, or removable storage media. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2134-2014 el 6 de agosto de 2014. Peligro reportado: The following defects have been identified for data loaded with Patient Data Manager version 2.0 (including subversions 2.0.0, 2.0.1 and 2.0.2): 1) When loading non-square pixel image data with PDM 2.0, the data set is not correctly transferred and displayed in the corresponding Brainlab navigation or planning software. When a data set is loaded, the pixels of the data set are sorted in order to
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter