FDAFDA · Alimentos y medicamentos·N.º oficial Z-2146-2014

ExacTrac is intended to be used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, in order to treat lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. ExacTrac may also be used to monitor the patient position during the treatment.

Publicado el 13 de agosto de 2014 · Inicio del retiro: 24 de junio de 2014

Resumen

FDA registra el retiro Z-2146-2014 de dispositivos médicos de Clase II, publicado el 13 de agosto de 2014 e iniciado el 24 de junio de 2014. Empresa responsable: Brainlab AG; Estado: Finalizado; Cantidad reportada: 15 systems; Lote: ExacTrac Version 6.0.4; Model/Catalog Numbers: 1) 20833C EXACTRAC 6.0 IR POSITIONING SOFTWARE; 2) 20834C EXACTRAC 6.0 X-RAY POSITIONING SOFTWARE; 3) 20835C EXACTRAC 6.0 IR MONITORING SOFTWARE; 4) 20853C EXACTRAC 6.0 X-RAY LIMITED SOFTWARE; 5) 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR); 6) 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR); 7) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR); 8) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED; 9) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC; 10) 49936A ET SOFTWARE UPDATE 6.0.X TO 6.0.4; 11) 49938 ET SOFTWARE UPDATE 6.0.3TO 6.0.4; 12) 49939 ET SOFTWARE UPDATE 6.0.X to 6.0.4 P&R; 13) 49973B ET DATA PREP / REVIEW SYSTEM; 14) 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY; 15) 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED; 16) 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

When using multiple isocenters (radiation treatment targets) within a single plan, in certain workflow conditions ExacTrac v.6.0.4 might move the patient to an unintended isocenter position, despite displaying the green "OK" icon. If this anomaly occurs and is not detected by the user, the radiation treatment dose at the linear accelerator may be delivered to the unintended target position.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Brainlab AG
Distribuidor
Worldwide Distribution - USA including CA, CT, and IL; Internationally to France, Germany, Japan, Netherlands, Russia, and Slovenia.
Lote
ExacTrac Version 6.0.4; Model/Catalog Numbers: 1) 20833C EXACTRAC 6.0 IR POSITIONING SOFTWARE; 2) 20834C EXACTRAC 6.0 X-RAY POSITIONING SOFTWARE; 3) 20835C EXACTRAC 6.0 IR MONITORING SOFTWARE; 4) 20853C EXACTRAC 6.0 X-RAY LIMITED SOFTWARE; 5) 49926B ET UPGRADE SOFTWARE 3.X TO 6.0 (IR+XR); 6) 49927B ET UPGRADE SOFTWARE 4.X TO 6.0 (IR+XR); 7) 49928B ET UPGRADE SOFTWARE 5.X TO 6.0 (IR+XR); 8) 49933B ET UPGRADE TRUEBEAM 5.5 TO 6.0 LIMITED; 9) 49934A EXACTRAC UPGRADE 5.5 TO 6.0 BASIC; 10) 49936A ET SOFTWARE UPDATE 6.0.X TO 6.0.4; 11) 49938 ET SOFTWARE UPDATE 6.0.3TO 6.0.4; 12) 49939 ET SOFTWARE UPDATE 6.0.X to 6.0.4 P&R; 13) 49973B ET DATA PREP / REVIEW SYSTEM; 14) 49996B ET UPGRADE SOFTWARE X.X TO 6.0 SW ONLY; 15) 49997B ET UPGRADE SOFTWARE 5.5 TO 6.0 LIMITED; 16) 49998B ET 6.0 DATA PREP/REVIEW SYSTEM UPG. KIT
Clasificación de riesgo
Class II
Área (FDA)
device
País
Germany

Retiros relacionados

Preguntas sobre este retiro

¿ExacTrac is intended to be used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, in order to treat lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. ExacTrac may also be used to monitor the patient position during the treatment. fue retirado del mercado?

Sí. FDA publicó el retiro Z-2146-2014 el 13 de agosto de 2014. Peligro reportado: When using multiple isocenters (radiation treatment targets) within a single plan, in certain workflow conditions ExacTrac v.6.0.4 might move the patient to an unintended isocenter position, despite displaying the green "OK" icon. If this anomaly occurs and is not detected by the user, the radiation treatment dose at the linear accelerator may be delivered to the unintended target position.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter