FDAFDA · Alimentos y medicamentos·N.º oficial Z-2153-2012

Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier: CP-08847; Model Number: HIT; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The ARIA Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Oncology also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments. ARIA also includes optional tools to manage the oncology department's clinical and business information.

Publicado el 15 de agosto de 2012 · Inicio del retiro: 2 de julio de 2012

Resumen

FDA registra el retiro Z-2153-2012 de dispositivos médicos de Clase II, publicado el 15 de agosto de 2012 e iniciado el 2 de julio de 2012. Empresa responsable: Varian Medical Systems, Inc. Oncology Systems; Estado: Finalizado; Cantidad reportada: 2 Units.; Lote: Serial Numbers/CODES: HIT0535, HIT2953. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

A software anomaly has been identified with the Import Export Application version 11.0 where MLC leaf positions for plans using BrainLab m3 and mMLC are incorrectly imported.

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Varian Medical Systems, Inc. Oncology Systems
Distribuidor
Nationwide Distribution including the states of New York and Ohio.
Lote
Serial Numbers/CODES: HIT0535, HIT2953
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿Varian brand Varian ARIA Oncology Information System, Import Export application, v 11.0; Reference/FSCA Identifier: CP-08847; Model Number: HIT; Product is manufactured and distributed by Varian Medical Systems Inc., Palo Alto, CA The ARIA Radiation Oncology product is a treatment plan and image management application. It enables the authorized user to enter, access, modify, store and archive treatment plan and image data from diagnostic studies, treatment planning, simulation, plan verification and treatment. ARIA Radiation Oncology also stores the treatment histories including dose delivered to defined sites and provides tools to verify performed treatments. ARIA also includes optional tools to manage the oncology department's clinical and business information. fue retirado del mercado?

Sí. FDA publicó el retiro Z-2153-2012 el 15 de agosto de 2012. Peligro reportado: A software anomaly has been identified with the Import Export Application version 11.0 where MLC leaf positions for plans using BrainLab m3 and mMLC are incorrectly imported.

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter