Fiber Cleaver; Fiber Cleaver is also used to score the outer jacket on EndoStat Fibers (Aura XP) The Fiber Cleaver is a reusable and sterilizable tool used to score the outer jacket on Endostat and SureFlex Fibers. The tool is a one piece pencil-like design with a sharpened blade at the end. The Fiber Cleaver is a SureFlex TM Reusable Laser Lithotripsy fiber and EndoStat fiber accessory. The Fiber Cleaver does not come in direct contact with the patient.
Publicado el 29 de julio de 2015 · Inicio del retiro: 28 de mayo de 2015
Resumen
FDA registra el retiro Z-2170-2015 de dispositivos médicos de Clase II, publicado el 29 de julio de 2015 e iniciado el 28 de mayo de 2015. Empresa responsable: American Medical Systems Innovation Center - Silicon Valley; Estado: Finalizado; Cantidad reportada: 16,710 total all affected devices; Lote: Model 0010-0760; All product manufactured from April 8, 2012 to April 8, 2015. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- American Medical Systems Innovation Center - Silicon Valley
- Distribuidor
- Worldwide Distribution - US Nationwide in the countries of: Argentina France Morocco, South Africa, Australia, Germany, Netherlands, Spain, Austria, Greece, New Zealand, Sri Lanka, Belgium, Guatemala, Norway, Switzerland, Bolivia, India, Panama, Thailand, Brazil, Ireland, Peru, Turkey, Canada, Israel, Poland, United Kingdom, Chile, Italy, Portugal, United States, Colombia, Korea, Qatar, Uruguay, Costa Rica, Lebanon, Romania, Finland, Malaysia and Saudi Arabia.
- Lote
- Model 0010-0760; All product manufactured from April 8, 2012 to April 8, 2015
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDATool, Stripper W/BLD, 18S, Nonautoclavable; The Fiber Stripper is a tool provided for stripping the protective29 de julio de 2015
- FDATool, Stripper W/BLD, 44S, Autoclavable; The Fiber Stripper is a tool provided for stripping the protective29 de julio de 2015
- FDATHOMAS by Gardner Denver, MEDI-PUMP(R) Aspirator, Model 1243 Product Usage: medical aspirator29 de julio de 2015
- FDAVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR Scout29 de julio de 2015
- FDAPerkin Elmer WIZARD2 1-detector,3", 1000 samples Product Code: 2480-0010 The Wizard®" gamma counter is29 de julio de 2015
- FDAARCHITECT c8000 System List Number 01G06-98 Product Usage: The Abbott ARCHITECT cSystem is intended for In29 de julio de 2015
Preguntas sobre este retiro
¿Fiber Cleaver; Fiber Cleaver is also used to score the outer jacket on EndoStat Fibers (Aura XP) The Fiber Cleaver is a reusable and sterilizable tool used to score the outer jacket on Endostat and SureFlex Fibers. The tool is a one piece pencil-like design with a sharpened blade at the end. The Fiber Cleaver is a SureFlex TM Reusable Laser Lithotripsy fiber and EndoStat fiber accessory. The Fiber Cleaver does not come in direct contact with the patient. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2170-2015 el 29 de julio de 2015. Peligro reportado: validation data related to cleaning instructions and sterilization methods and the methods defined in instructions for use (IFU) for the Sureflex Reusable Fibers and the Reusable Stripper and Cleaver Accessories do not meet the new guidance on qualifying reusable products
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter