FDAFDA · Alimentos y medicamentos·N.º oficial Z-2181-2012
Kimberly Clark Patient Warming System - Model 1000 Control Unit. Kimberly Clark Patient Warming System - Model 1000 Control Unit and Energy Transfer Pads is a thermal regulating system, indicated for monitoring and controlling patient temperature.
Publicado el 15 de agosto de 2012 · Inicio del retiro: 27 de abril de 2012
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
A defect of the "fuse holder" can lead to a Control Unit failure that may cause the device to lose its ability to warm and maintain set temperature and may also result in excessive heat with the fuse assembly that could lead to the generation of smoke.
Qué hacer
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Detalles del retiro
- Fabricante
- Kimberly-Clark Corporation
- Distribuidor
- Nationwide Distribution-including the states of AZ, CA, CO, DE, FL, GA, IA, ID, IN, IL, KS, KY, LA, MA, MD, ME, MO, MS, NE, NM, OH, OR, PA, RI, SC, TN, TX, VA, WI and WV.
- Lote
- Catalogue #01000-01 and 01000-01R, Units with 4-Digit Serial Numbers: Lot #4297, 4324, 4332, 4519, 4559, 4572; Units with 10 Character Serial Numbers: SA08017027, SA08017028, SA08017029, SA08017030, SA08017031, SA08017032, SA08017033, SA08017034, SA08017035, SA08017036, SA08017037, SA08017038, SA08017039, SA08017040, SA08017041, SA08017042, SA08017043, SA08017044, SA08017045, SA08017046, SA08017047, SA08017048, SA08017049, SA08343100, SA11137004, SA11137005, SA11137006, SA11137007, SA11137008, SA08094124, SA08094125, SA08094126, SA08094127, SA08094128, SA08094129, SA08094130, SA08094131, SA08094132, SA08094133, SA08094134, SA08094135, SA08094136, SA08094137, SA08094138, SA08094139, SA08094140, SA08094141, SA08094142, SA08094143, SA08094144, SA08094145, SA08094146, SA08094147, SA08094148, SA08094149, SA08094150, SA08094151, SA08094152, SA08094153, SA09020004, SA09020005, SA09020007, SA09020008, SA09020010, SA09020013, SA11298011, SA11298012, SA11298013, SA11298014, SA11298015, SA11298016, SA11298017, SA11298018, SA11298019, SA08136001, SA08136002, SA08136003, SA08136004, SA08136005, SA08136006, SA08136007, SA08136008, SA08136009, SA08136010, SA09043001, SA09043002, SA09043003, SA09043004, SA09043007, SA09043008, SA09043009, SA09043010, SA09043010, SA09043011, SA09043012, SA09043013, SA09043014, SA09043015, SA09043016, SA09043017, SA09043018, SA09043019, SA09043020, SA08343061, SA08343082, SA08343083, SA08343084, SA08343085, SA08343086, SA08343087, SA08343088, SA08343089, SA08343090, SA08343091, SA08343092, SA08343093, SA08343094, SA08343095, SA08343096, SA10337006, SA10337007, SA10337008, SA10337009 and SA10337010.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device