Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2211-2012

VITROS Chemistry Products DGXN Slides --- [REF] 834 3386. Product is packaged in a box with multiple slides. The lot and expiration codes are found on the side of the package. Product Usage: Intended use: For in vitro diagnostic use only. VITROS Chemistry Products DGXN Slides quantitatively measure digoxin (DGXN) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System.

Publicado el 22 de agosto de 2012 · Inicio del retiro: 28 de junio de 2012

Resumen

FDA registra el retiro Z-2211-2012 de dispositivos médicos de Clase II, publicado el 22 de agosto de 2012 e iniciado el 28 de junio de 2012. Empresa responsable: Ortho-Clinical Diagnostics; Estado: Finalizado; Cantidad reportada: 13338 sale units; Lote: Lots from coating 0204, 0205, 0206. A Coating (CTG) is a unique identifier within the Lot Number of some of OCDs products. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Ortho Clinical Diagnostics revised the VITROS Chemistry Products DGXN Slides Instructions for Use (IFU) and removed heparin plasma as a recommended specimen type for VITROS DGXN Slides

Qué hacer

Voluntary: Firm initiated · Letter

Recibe un correo cuando FDA publique un retiro como este

Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.

Detalles del retiro

Fabricante
Ortho-Clinical Diagnostics
Distribuidor
Worldwide Distribution - US (nationwide)
Lote
Lots from coating 0204, 0205, 0206. A Coating (CTG) is a unique identifier within the Lot Number of some of OCDs products
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

Retiros relacionados

Preguntas sobre este retiro

¿VITROS Chemistry Products DGXN Slides --- [REF] 834 3386. Product is packaged in a box with multiple slides. The lot and expiration codes are found on the side of the package. Product Usage: Intended use: For in vitro diagnostic use only. VITROS Chemistry Products DGXN Slides quantitatively measure digoxin (DGXN) concentration in serum and plasma using VITROS 250/350/950/5,1 FS and 4600 Chemistry Systems and the VITROS 5600 Integrated System. fue retirado del mercado?

Sí. FDA publicó el retiro Z-2211-2012 el 22 de agosto de 2012. Peligro reportado: Ortho Clinical Diagnostics revised the VITROS Chemistry Products DGXN Slides Instructions for Use (IFU) and removed heparin plasma as a recommended specimen type for VITROS DGXN Slides

¿Qué debo hacer si tengo este producto?

Voluntary: Firm initiated · Letter