Large Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T, DS50-103-XL-T
Publicado el 14 de junio de 2017 · Inicio del retiro: 26 de abril de 2017
Resumen
FDA registra el retiro Z-2220-2017 de dispositivos médicos de Clase II, publicado el 14 de junio de 2017 e iniciado el 26 de abril de 2017. Empresa responsable: Iba Dosimetry; Estado: Finalizado; Cantidad reportada: 110 units; Lote: DS50-001-XL-T, Serial Numbers: 24887, 25084, 25085 DS50-002-XL-T, Serial Numbers: 24185, 25219, 25470, 24073, 24187, 24480, 24888, 25089, 24889, 24184, 24813, 24814, 24866, 24867, 24890, 25088, 25220, 24482, 24186, 25090, 24626, 24074, 24189, 25218, 25468, 25469 DS50-003-XL-T, Serial Numbers: 24437, 24260, 24639, 25472, 24257, 24291, 24630, 24640, 24868, 25093, 25478, 24180, 24182, 24259, 24323, 24439, 25222, 24179, 24290, 24293, 24633, 24637, 25091, 25223, 24627, 24077, 24634, 24438, 25476, 25221, 24292, 24263, 24322, 24435, 24628, 24642, 25094, 25479, 24177, 24178, 24641, 25481, 24261, 24076, 24871, 25224, 24078, 24079, 24080, 24175, 24176, 24183, 24256, 24258, 24262, 24264, 24265, 24289, 24321, 24436, 24629, 24631, 24635, 24636, 24638, 24869, 24891, 24892, 24893, 24894, 24896, 24897, 25092, 25095, 25225, 25477 DS50-102-XL-T, Serial Numbers: 24190, 24481, 25467 DS50-103-XL-T, Serial Numbers: 24870, 24898. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Iba Dosimetry
- Distribuidor
- Worldwide Distribution - USA (nationwide) and to the countries of : Argentina, Australia, Austria, Canada, Cyprus, France, Germany, Hong Kong, Ireland, Islamic Republic of, Israel, Italy, Japan, Kingdom of Saudi Arabia, Kuwait, New Zealand, People's Republic of China, Philippines, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, R.O.C., and United Kingdom.
- Lote
- DS50-001-XL-T, Serial Numbers: 24887, 25084, 25085 DS50-002-XL-T, Serial Numbers: 24185, 25219, 25470, 24073, 24187, 24480, 24888, 25089, 24889, 24184, 24813, 24814, 24866, 24867, 24890, 25088, 25220, 24482, 24186, 25090, 24626, 24074, 24189, 25218, 25468, 25469 DS50-003-XL-T, Serial Numbers: 24437, 24260, 24639, 25472, 24257, 24291, 24630, 24640, 24868, 25093, 25478, 24180, 24182, 24259, 24323, 24439, 25222, 24179, 24290, 24293, 24633, 24637, 25091, 25223, 24627, 24077, 24634, 24438, 25476, 25221, 24292, 24263, 24322, 24435, 24628, 24642, 25094, 25479, 24177, 24178, 24641, 25481, 24261, 24076, 24871, 25224, 24078, 24079, 24080, 24175, 24176, 24183, 24256, 24258, 24262, 24264, 24265, 24289, 24321, 24436, 24629, 24631, 24635, 24636, 24638, 24869, 24891, 24892, 24893, 24894, 24896, 24897, 25092, 25095, 25225, 25477 DS50-102-XL-T, Serial Numbers: 24190, 24481, 25467 DS50-103-XL-T, Serial Numbers: 24870, 24898
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAEllex Tango Reflex with slit lamp Laser Ophthalmic In the SLT mode, the device is intended to be used for14 de junio de 2017
- FDAUrea Base Agar, Acumedia PN 7226 Urea Agar Base is used with agar for the differentiation of microorganisms on14 de junio de 2017
- FDARavioli Orso. Product is made refrigerated. Product is packaged in plastic containers. The plastic containers14 de junio de 2017
- FDAMagellan Diagnostics LeadCare and LeadCare II test Kits. Product Usage: The LeadCare Blood Lead Test Kit is14 de junio de 2017
- FDAPapperdelli ¿ inch. Product is made refrigerated. Product is packaged in plastic containers. The plastic14 de junio de 2017
- FDASesamoid Plasty, Stereotaxic Instrument The Sesamoid Plasty workstation is used to run the tracking software14 de junio de 2017
Preguntas sobre este retiro
¿Large Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T, DS50-103-XL-T fue retirado del mercado?
Sí. FDA publicó el retiro Z-2220-2017 el 14 de junio de 2017. Peligro reportado: When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter