FDAFDA · Alimentos y medicamentos·N.º oficial Z-2220-2017
Large Stealth Chamber, Catalog Numbers: DS50-001-XL-T, DS50-002-XL-T, DS50-003-XL-T, DS50-102-XL-T, DS50-103-XL-T
Publicado el 14 de junio de 2017 · Inicio del retiro: 26 de abril de 2017
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
When the Stealth Chamber is used as a reference detector in relative dosimetry, deviation in the measured PDDs at shallow depths and geometric shadowing of measured profiles may appear.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- Iba Dosimetry
- Distribuidor
- Worldwide Distribution - USA (nationwide) and to the countries of : Argentina, Australia, Austria, Canada, Cyprus, France, Germany, Hong Kong, Ireland, Islamic Republic of, Israel, Italy, Japan, Kingdom of Saudi Arabia, Kuwait, New Zealand, People's Republic of China, Philippines, Russia, Slovakia, South Africa, South Korea, Spain, Sweden, Taiwan, R.O.C., and United Kingdom.
- Lote
- DS50-001-XL-T, Serial Numbers: 24887, 25084, 25085 DS50-002-XL-T, Serial Numbers: 24185, 25219, 25470, 24073, 24187, 24480, 24888, 25089, 24889, 24184, 24813, 24814, 24866, 24867, 24890, 25088, 25220, 24482, 24186, 25090, 24626, 24074, 24189, 25218, 25468, 25469 DS50-003-XL-T, Serial Numbers: 24437, 24260, 24639, 25472, 24257, 24291, 24630, 24640, 24868, 25093, 25478, 24180, 24182, 24259, 24323, 24439, 25222, 24179, 24290, 24293, 24633, 24637, 25091, 25223, 24627, 24077, 24634, 24438, 25476, 25221, 24292, 24263, 24322, 24435, 24628, 24642, 25094, 25479, 24177, 24178, 24641, 25481, 24261, 24076, 24871, 25224, 24078, 24079, 24080, 24175, 24176, 24183, 24256, 24258, 24262, 24264, 24265, 24289, 24321, 24436, 24629, 24631, 24635, 24636, 24638, 24869, 24891, 24892, 24893, 24894, 24896, 24897, 25092, 25095, 25225, 25477 DS50-102-XL-T, Serial Numbers: 24190, 24481, 25467 DS50-103-XL-T, Serial Numbers: 24870, 24898
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States