FDAFDA · Alimentos y medicamentos·N.º oficial Z-2228-2018
Part #: SAM XT-C, Model: Hi-Viz Blue
Publicado el 20 de junio de 2018 · Inicio del retiro: 2 de mayo de 2018
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Based on internal testing, results indicated a possible failure of the stitches securing the buckle to the nylon belt could occur, posing a potential risk when used on a human patient to stop arterial blood flow. There is a remote chance that a serious injury and/or death could occur as a result of the failure of the device.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- The Seaberg Company Inc
- Distribuidor
- USA: CA, MD, OH, FL, ID, CO, CT, OR, AZ, IL, ME, MO, NY, SC, MI, AR, MT, CT, WA, TX, KY, NC, NJ, PA OUS: Australia, Canada, China, Columbia, Costa Rica, Europe (EU), Hong Kong, Iceland, India, Iraq, Israel, Japan, Kenya, Korea, Lebanon, Malaysia, Russia, Saudi Arabia, Singapore, South Africa, Thailand, UAE
- Lote
- Lot #: XT1808, XT1809, XT1810
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States