Volver al buscador
FDAFDA · Alimentos y medicamentos·N.º oficial Z-2263-2013

Reliance Endoscope Processing System (REPS), STERIS Corporation, 5960 Heisley Road, Mentor, OH 44060-1834 USA www.steris.com. The intended use of the REPS is washing and high level disinfection of up to two manually pre-cleaned, immersible, reusable, heat-resistant, semi-critical devices.

Publicado el 2 de octubre de 2013 · Inicio del retiro: 29 de agosto de 2013

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

The air pipe that delivers unfiltered ambient air from the procedure room into the water bottle on the listed FUJINON G5 & G8 type flexible GI endoscopes does not connect to the main air/water channel within the light guide (LG) connection head and, as a result, the air pipe does not receive forced flow during the Reliance EPS processing cycle.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Steris Corporation
Distribuidor
Worldwide Distribution-USA (nationwide) including the states of AK, CO, FL, GA, HI, ID, IL, KS, LA, MD, MN, MO, MT, NE, NM, OH, PA, SC, UT, & WY and the countries of Australia, Canada & Spain.
Lote
Model #'s: MB30001, MB30002, MB30003, MB000077, MB000078, MB000079, MB000080 and MB000081. Serial number range 3625606001 - 3608912002
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States

¿Vendes productos como este?

Recalert vigila todo tu catálogo contra retiros como este, automáticamente.

Empezar gratis — sin tarjeta