FDAFDA · Alimentos y medicamentos·N.º oficial Z-2283-2015
ANSPACH - Fluted Drum, Side Cut only, 6MM x 14MM REF L-6DRC-G1 Cutting shaping bone including spine and cranium.
Publicado el 5 de agosto de 2015 · Inicio del retiro: 13 de abril de 2015
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
G1 Dissection Tools are not compatible with all attachments listed on the labeling; G1 cutting tools are labeled with a Specified Shaft Exposure which may not be consistent when used with each attachment on the product label. Shaft exposure may be greater.
Qué hacer
Voluntary: Firm initiated · Letter
Detalles del retiro
- Fabricante
- The Anspach Effort, Inc.
- Distribuidor
- Worldwide Distribution - US including AZ, CA, CO, FL, GA, IL, KY, LA, MN, MO, NC, NJ, NY, OH, OR, PA, TX, and Internationally to Canada and China.
- Lote
- REF L-6DRC-G1 ; ALL lot numbers
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
¿Vendes productos como este?