FDAFDA · Alimentos y medicamentos·N.º oficial Z-229-2014
HeartMate II LVAD Pump and Pock Controller Kit; Product Usage: HeartMate II is a mechanical device that circulates blood throughout the body when the heart is too weak to pump blood adequately on its own.
Publicado el 2 de abril de 2014 · Inicio del retiro: 4 de marzo de 2014
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Discovery of serious injuries and deaths associated with the process of changing from a primary System controller to their back-up System controller in patients using the Pocket System controller model.
Qué hacer
Voluntary: Firm initiated · Press Release
Detalles del retiro
- Fabricante
- Thoratec Corporation
- Distribuidor
- Worldwide distribution: US (nationwide) including states of: AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI; and countries of: Austria, Belgium, Czech Republic, France, Germany, Ireland, Israel, Italy, Lebanon, Lithuania, Poland, Saudi Arabia, Singapore, Slovenia, Sweden, Switzerland, The Netherlands, Turkey, and United Kingdom.
- Lote
- Catalog number 107801 all serial numbers
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States