Capintec Captus 3000 Thyroid Uptake System Nuclear uptake probe intended to be used by trained Nuclear Medicine Technologists to perform thyroid uptake procedures, which consist of the measurement of organ uptake and circulating radioactivity.
Publicado el 12 de septiembre de 2012 · Inicio del retiro: 24 de julio de 2012
Resumen
FDA registra el retiro Z-2312-2012 de dispositivos médicos de Clase II, publicado el 12 de septiembre de 2012 e iniciado el 24 de julio de 2012. Empresa responsable: Capintec Inc; Estado: Finalizado; Cantidad reportada: 256; Lote: Catalog numbers for new systems: 5430-0076, 5430-0077, 5430-0103, and 5430-0087. Catalog numbers for conversion systems: 5430-0118 and 5430-0119. Serial numbers for new systems: 901006 through 901394, inclusive. Serial numbers for conversion systems: CNV 674 through CNV 730, inclusive. Also affected are devices which were serviced and devices for which customers received software upgrades between January 20, 2011 and June 30, 2012 with Revision 1.22, 1.23, 1.24, 1.25, or 1.26. (NOTE: Item number 5430-0075, Captus 3000 well counter system, is excluded. This system does not have the ability to perform thyroid uptake procedures.). La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Customers who received a Capintec Captus 3000 Thyroid Uptake System with Software Revisions 1.22, 1.23, 1.24, 1.25, and 1.26 may encounter an error while using the thyroid uptake protocol ?Measure Liquid and Subtract Residual." This protocol was a new feature added to Software Revision 1.22 in January 2011. The error occurs only if the user does NOT measure the residual dose immediately after me
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Capintec Inc
- Distribuidor
- Worldwide distribution: USA (nationwide) and countries of: Argentina, Belgium, Brazil, Canada, China, Greece, Hong Kong, India, Israel, Malaysia, Philippines, Spain, South Korea, Thailand, and Turkey.
- Lote
- Catalog numbers for new systems: 5430-0076, 5430-0077, 5430-0103, and 5430-0087. Catalog numbers for conversion systems: 5430-0118 and 5430-0119. Serial numbers for new systems: 901006 through 901394, inclusive. Serial numbers for conversion systems: CNV 674 through CNV 730, inclusive. Also affected are devices which were serviced and devices for which customers received software upgrades between January 20, 2011 and June 30, 2012 with Revision 1.22, 1.23, 1.24, 1.25, or 1.26. (NOTE: Item number 5430-0075, Captus 3000 well counter system, is excluded. This system does not have the ability to perform thyroid uptake procedures.)
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACaptus 4000e Thyroid Uptake System, Model Numbers 5430-30151 and 5430-30152 - Product Usage: is intended to be1 de abril de 2020
- FDAAcrobat 3000 (AC3000) Spring Arm as support arm for iLED operating lights, TRUMPF Medizinsysteme GmbH + Co KG12 de septiembre de 2012
- FDASiemens ADVIA Centaur Multi-Diluent 1 - Used for automated dilutions for the ADVIA Centaur BNP and TSH3 Ultra12 de septiembre de 2012
- FDAOPTETRAK LOGIC TIBIAL INSERT POSTERIOR STABILIZED***Match Component Colors 6 YELLOW, 11mm, Use with LOGIC12 de septiembre de 2012
- FDATxRx Knee 15 Coil 3T. For use in conjunction with magnetic resonance scanners to produce diagnostic images of12 de septiembre de 2012
- FDAZilver 518 Vascular Self Expanding Stent The Zilver Vascular Stent is intended for use as an adjunct to12 de septiembre de 2012
Preguntas sobre este retiro
¿Capintec Captus 3000 Thyroid Uptake System Nuclear uptake probe intended to be used by trained Nuclear Medicine Technologists to perform thyroid uptake procedures, which consist of the measurement of organ uptake and circulating radioactivity. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2312-2012 el 12 de septiembre de 2012. Peligro reportado: Customers who received a Capintec Captus 3000 Thyroid Uptake System with Software Revisions 1.22, 1.23, 1.24, 1.25, and 1.26 may encounter an error while using the thyroid uptake protocol ?Measure Liquid and Subtract Residual." This protocol was a new feature added to Software Revision 1.22 in January 2011. The error occurs only if the user does NOT measure the residual dose immediately after me
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter