MNA-420 ENC HDMI, Material Number K9303311; cathode ray tube display

Publicado el 17 de julio de 2024 · Inicio del retiro: 24 de abril de 2024

Estado reportado al momento de publicación: Completed — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

We have identified a potential issue with the SFP+ module supplied by Molex (Part Number: B620050) when connected to the ETH0 slot of either the MNA-6x0 ENC HDMI or MNA-4x0 ENC HDMI. In some instances, this connection may render the first HDMI channel inaccessible, making both the channel and the respective MNA ENC unit unusable. Additionally, readings from the module become unavailable when this issue arises. Similarly, when this Molex module is connected to the ETH0 slot of an MNA-620 ENC ANA, readings may also become inaccessible.

Qué hacer

Voluntary: Firm initiated · E-Mail

Detalles del retiro

Fabricante
Barco N.V.
Distribuidor
US: FL
Lote
serial numbers: 2530986299, 2530986300, 2530986301, 2530986302, 2530986303, 2530986304, 2530986305, 2530986306, 2530986307, 2530986308, 2530986309, 2530986310, 2530986311, 2530986312, 2530986313, 2530986314, 2530986315, 2530986316, 2530986317, 2530986318, 2530986319, 2530986320, 2530986321, 2530986322, 2530986323, 2530986324, 2530986325, 2530986326, 2530986327, 2531041594, 2531041591, 2531041578, 2531041585, 2531041595, 2531041596, 2531041586, 2531041588, 2531041587, 2531041577, 2531041584, 2531041581, 2531041612, 2531041608, 2531041600, 2531041601, 2531041604, 2531041599, 2531041602, 2531041605, 2531041606, 2531041607, 2531041609, 2531041610, 2531041580, 2531041583, 2531041598, 2531041593, 2531041590, 2531041614, 2531041582, 2531041613, 2531041589, 2531041616, 2531041579, 2531041615, 2531041603, 2531041597, 2531041592, 2530985205, 2530986280, 2530986281, 2530986282, 2530986283, 2530986284, 2530986285, 2530986286, 2530986287, 2530986288, 2530986289, 2530986290, 2530986291, 2530986292, 2530986293, 2530986294, 2530986295, 2530986296, 2530986297, 2530986298
Clasificación de riesgo
Class II
Área (FDA)
device
País
Belgium