Ultraview SL 91493 Integrated Module Housing. PCIS patient monitors is intended to sound various alarm tones that are generated by the specific PCIS patient parameter modules. Each patient parameter module determines (a) when an alarm condition is violated; (b) the alarm priority (i.e. high, medium or low); (c) alarm limits; and (d) when to generate and terminate alarms. PCIS patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. Spacelabs Medical PCIS patient monitors are designed to communicate with a variety of external devices such as display devices, network devices, serial devices, user input devices, audio system, local/remote recorders and Spacelabs Medical patient parameter modules. In all applications, Spacelabs Medical PCIS patient monitors provide access to that data on the Spacelabs Medical network either via hardwire or wireless transmission. Through this network, a patient's data may be displayed at the bedside monitor, central station or clinical information system.
Publicado el 26 de septiembre de 2012 · Inicio del retiro: 27 de agosto de 2012
Resumen
FDA registra el retiro Z-2382-2012 de dispositivos médicos de Clase II, publicado el 26 de septiembre de 2012 e iniciado el 27 de agosto de 2012. Empresa responsable: Spacelabs Healthcare, Llc; Estado: Finalizado; Cantidad reportada: 527 units total; 350 US and 177 International; Lote: Domestic - 1493-000275, 1493-000276, 1493-000277, 1493-002469, 1493-002616, 1493-003358, 1493-003359, 1493-003667, 1493-003668, 1493-003669, 1493-003670, 1493-003671, 1493-003672, 1493-003673, 1493-003674, 1493-003675, 1493-003676, 1493-003677, 1493-003678, 1493-003679, 1493-003680, 1493-003681, 1493-003682, 1493-003683, 1493-003684, 1493-003685, 1493-003686, 1493-003687, 1493-003688, 1493-003693, 1493-003694, 1493-003695, 1493-003696, 1493-003697, 1493-003879, 1493-003880, 1493-003881, 1493-003882, 1493-003883, 1493-003884, 1493-003885, 1493-003886, 1493-003887, 1493-003888, 1493-003889, 1493-003890, 1493-003891, 1493-003892, 1493-003893, 1493-003894, 1493-003895, 1493-003896, 1493-003897, 1493-003898, 1493-003899, 1493-003900, 1493-003901, 1493-003902, 1493-003903, 1493-003904, 1493-003905, 1493-003906, 1493-003907, 1493-003908, 1493-003909, 1493-003914, 1493-003915, 1493-003916, 1493-003917, 1493-003918, 1493-003919, 1493-003920, 1493-003921, 1493-003922, 1493-003923, 1493-003924, 1493-003925, 1493-003926, 1493-003927, 1493-003928, 1493-003929, 1493-003930, 1493-003931, 1493-003932, 1493-003933, 1493-003934, 1493-003935, 1493-003936, 1493-003937, 1493-003938, 1493-003939, 1493-003940, 1493-003941, 1493-003942, 1493-003943, 1493-003944, 1493-003945, 1493-003957, 1493-003958, 1493-003959, 1493-003960, 1493-003961, 1493-003962, 1493-003963, 1493-003964, 1493-003965, 1493-003985, 1493-004007, 1493-004011, 1493-004023, 1493-004026, 1493-004027, 1493-004028, 1493-004029, 1493-004030, 1493-004031, 1493-004032, 1493-004033, 1493-004034, 1493-004043, 1493-004044, 1493-004047, 1493-004048, 1493-004050, 1493-004051, 1493-004052, 1493-004053, 1493-004054, 1493-004055, 1493-004056, 1493-004058, 1493-004060, 1493-004064, 1493-004067, 1493-004068, 1493-004069, 1493-004070, 1493-004071, 1493-004075, 1493-004084, 1493-004085, 1493-004086, 1493-004106, 1493-004112, 1493-004113, 1493-004114, 1493-004115, 1493-004116, 1493-004117, 1493-004118, 1493-004119, 1493-004120, 1493-004121, 1493-004122, 1493-004123, 1493-004124, 1493-004125, 1493-004126, 1493-004127, 1493-004128, 1493-004130, 1493-004158, 1493-004159, 1493-004160, 1493-004161, 1493-004162, 1493-004163, 1493-004164, 1493-004172, 1493-004173, 1493-004174, 1493-004175, 1493-004176, 1493-004177, 1493-004178, 1493-004179, 1493-004180, 1493-004181, 1493-004182, 1493-004183, 1493-004184, 1493-004185, 1493-004205, 1493-004206, 1493-004207, 1493-004208, 1493-004209, 1493-004210, 1493-004211, 1493-004212, 1493-004214, 1493-004215, 1493-004216, 1493-004217, 1493-004218, 1493-004219, 1493-004220, 1493-004221, 1493-004222, 1493-004223, 1493-004224, 1493-004225, 1493-004226, 1493-004227, 1493-004228, 1493-004229, 1493-004230, 1493-004231, 1493-004232, 1493-004233, 1493-004234, 1493-004235, 1493-004236, 1493-004237, 1493-004238, 1493-004239, 1493-004240, 1493-004241, 1493-004242, 1493-004243, 1493-004256, 1493-004257, 1493-004258, 1493-004259, 1493-004260, 1493-004261, 1493-004262, 1493-004263, 1493-004264, 1493-004265, 1493-004266, 1493-004267, 1493-004268, 1493-004269, 1493-004270, 1493-004271, 1493-004272, 1493-004273, 1493-004274, 1493-004275, 1493-004276, 1493-004277, 1493-004278, 1493-004279, 1493-004280, 1493-004281, 1493-004282, 1493-004283, 1493-004284, 1493-004285, 1493-004286, 1493-004287, 1493-004288, 1493-004289, 1493-004290, 1493-004291, 1493-004292, 1493-004293, 1493-004294, 1493-004295, 1493-004296, 1493-004297, 1493-004298, 1493-004299, 1493-004301, 1493-004302, 1493-004306, 1493-004307, 1493-004308, 1493-004309, 1493-004310, 1493-004311, 1493-004312, 1493-004313, 1493-004314, 1493-004315, 1493-004316, 1493-004317, 1493-004318, 1493-004319, 1493-004320, 1493-004321, 1493-004322, 1493-004323, 1493-004324, 1493-004325, 1493-004326, 1493-004327, 1493-004328, 1493-004329, 1493-004330, 1493-004331, 1493-004332, 1493-004333, 1493-004334, 1493-004335, 1493-004336, 1493-004337, 1493-004338, 1493-004339, 1493-004340, 1493-004341, 1493-004342, 1493-004343, 1493-004344, 1493-004345, 1493-004346. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
There is a potential failure of the Integrated Module Housing causing the patient module parameters to drop off the monitor display or the patient module will fail to sign on when inserted into the housing.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Spacelabs Healthcare, Llc
- Distribuidor
- Worldwide Distribution.
- Lote
- Domestic - 1493-000275, 1493-000276, 1493-000277, 1493-002469, 1493-002616, 1493-003358, 1493-003359, 1493-003667, 1493-003668, 1493-003669, 1493-003670, 1493-003671, 1493-003672, 1493-003673, 1493-003674, 1493-003675, 1493-003676, 1493-003677, 1493-003678, 1493-003679, 1493-003680, 1493-003681, 1493-003682, 1493-003683, 1493-003684, 1493-003685, 1493-003686, 1493-003687, 1493-003688, 1493-003693, 1493-003694, 1493-003695, 1493-003696, 1493-003697, 1493-003879, 1493-003880, 1493-003881, 1493-003882, 1493-003883, 1493-003884, 1493-003885, 1493-003886, 1493-003887, 1493-003888, 1493-003889, 1493-003890, 1493-003891, 1493-003892, 1493-003893, 1493-003894, 1493-003895, 1493-003896, 1493-003897, 1493-003898, 1493-003899, 1493-003900, 1493-003901, 1493-003902, 1493-003903, 1493-003904, 1493-003905, 1493-003906, 1493-003907, 1493-003908, 1493-003909, 1493-003914, 1493-003915, 1493-003916, 1493-003917, 1493-003918, 1493-003919, 1493-003920, 1493-003921, 1493-003922, 1493-003923, 1493-003924, 1493-003925, 1493-003926, 1493-003927, 1493-003928, 1493-003929, 1493-003930, 1493-003931, 1493-003932, 1493-003933, 1493-003934, 1493-003935, 1493-003936, 1493-003937, 1493-003938, 1493-003939, 1493-003940, 1493-003941, 1493-003942, 1493-003943, 1493-003944, 1493-003945, 1493-003957, 1493-003958, 1493-003959, 1493-003960, 1493-003961, 1493-003962, 1493-003963, 1493-003964, 1493-003965, 1493-003985, 1493-004007, 1493-004011, 1493-004023, 1493-004026, 1493-004027, 1493-004028, 1493-004029, 1493-004030, 1493-004031, 1493-004032, 1493-004033, 1493-004034, 1493-004043, 1493-004044, 1493-004047, 1493-004048, 1493-004050, 1493-004051, 1493-004052, 1493-004053, 1493-004054, 1493-004055, 1493-004056, 1493-004058, 1493-004060, 1493-004064, 1493-004067, 1493-004068, 1493-004069, 1493-004070, 1493-004071, 1493-004075, 1493-004084, 1493-004085, 1493-004086, 1493-004106, 1493-004112, 1493-004113, 1493-004114, 1493-004115, 1493-004116, 1493-004117, 1493-004118, 1493-004119, 1493-004120, 1493-004121, 1493-004122, 1493-004123, 1493-004124, 1493-004125, 1493-004126, 1493-004127, 1493-004128, 1493-004130, 1493-004158, 1493-004159, 1493-004160, 1493-004161, 1493-004162, 1493-004163, 1493-004164, 1493-004172, 1493-004173, 1493-004174, 1493-004175, 1493-004176, 1493-004177, 1493-004178, 1493-004179, 1493-004180, 1493-004181, 1493-004182, 1493-004183, 1493-004184, 1493-004185, 1493-004205, 1493-004206, 1493-004207, 1493-004208, 1493-004209, 1493-004210, 1493-004211, 1493-004212, 1493-004214, 1493-004215, 1493-004216, 1493-004217, 1493-004218, 1493-004219, 1493-004220, 1493-004221, 1493-004222, 1493-004223, 1493-004224, 1493-004225, 1493-004226, 1493-004227, 1493-004228, 1493-004229, 1493-004230, 1493-004231, 1493-004232, 1493-004233, 1493-004234, 1493-004235, 1493-004236, 1493-004237, 1493-004238, 1493-004239, 1493-004240, 1493-004241, 1493-004242, 1493-004243, 1493-004256, 1493-004257, 1493-004258, 1493-004259, 1493-004260, 1493-004261, 1493-004262, 1493-004263, 1493-004264, 1493-004265, 1493-004266, 1493-004267, 1493-004268, 1493-004269, 1493-004270, 1493-004271, 1493-004272, 1493-004273, 1493-004274, 1493-004275, 1493-004276, 1493-004277, 1493-004278, 1493-004279, 1493-004280, 1493-004281, 1493-004282, 1493-004283, 1493-004284, 1493-004285, 1493-004286, 1493-004287, 1493-004288, 1493-004289, 1493-004290, 1493-004291, 1493-004292, 1493-004293, 1493-004294, 1493-004295, 1493-004296, 1493-004297, 1493-004298, 1493-004299, 1493-004301, 1493-004302, 1493-004306, 1493-004307, 1493-004308, 1493-004309, 1493-004310, 1493-004311, 1493-004312, 1493-004313, 1493-004314, 1493-004315, 1493-004316, 1493-004317, 1493-004318, 1493-004319, 1493-004320, 1493-004321, 1493-004322, 1493-004323, 1493-004324, 1493-004325, 1493-004326, 1493-004327, 1493-004328, 1493-004329, 1493-004330, 1493-004331, 1493-004332, 1493-004333, 1493-004334, 1493-004335, 1493-004336, 1493-004337, 1493-004338, 1493-004339, 1493-004340, 1493-004341, 1493-004342, 1493-004343, 1493-004344, 1493-004345, 1493-004346
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDASpacelabs Healthcare G2 Clinical Access Mislabeled CDs for PN: 063-1829-11 Rev A and Update Kit, PN:2 de abril de 2014
- FDAARIATELE TELEMETRY TRANSMITTER, Model 96281, with SpO2 Option C. Model 96281 is intended for use with either5 de febrero de 2014
- FDASpacelabs Healthcare qube Compact Patient Monitor, Model 91390. It is a patient monitor with arrhythmia27 de noviembre de 2013
- FDASpacelabs Healthcare elance Vital Signs Monitor, Model 93300, with Option S. It is a patient monitor with20 de noviembre de 2013
- FDAThe 2.7mm Drilling Guide is a reusable instrument used in conjunction with a drill to prepare oriented holes26 de septiembre de 2012
- FDAVarian brand Clinac, Trilogy, Novalis Tx, Unique, linear accelerators; Model Number: H14, H18, H29,26 de septiembre de 2012
Preguntas sobre este retiro
¿Ultraview SL 91493 Integrated Module Housing. PCIS patient monitors is intended to sound various alarm tones that are generated by the specific PCIS patient parameter modules. Each patient parameter module determines (a) when an alarm condition is violated; (b) the alarm priority (i.e. high, medium or low); (c) alarm limits; and (d) when to generate and terminate alarms. PCIS patient monitors are also capable of displaying alarm conditions on other monitors that are on the network through the Alarm Watch feature. Spacelabs Medical PCIS patient monitors are designed to communicate with a variety of external devices such as display devices, network devices, serial devices, user input devices, audio system, local/remote recorders and Spacelabs Medical patient parameter modules. In all applications, Spacelabs Medical PCIS patient monitors provide access to that data on the Spacelabs Medical network either via hardwire or wireless transmission. Through this network, a patient's data may be displayed at the bedside monitor, central station or clinical information system. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2382-2012 el 26 de septiembre de 2012. Peligro reportado: There is a potential failure of the Integrated Module Housing causing the patient module parameters to drop off the monitor display or the patient module will fail to sign on when inserted into the housing.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter