The 2.7mm Drilling Guide is a reusable instrument used in conjunction with a drill to prepare oriented holes in bone for insertion of a Standard Surfix¿ Locking Screw with plates for the Tiblaxis, Uni-CP, Advansys MLP and Advansys DLP compression plate systems. The Drilling Guides are initially supplied to users in kits containing the devices that comprise the Tibiaxis, Uni-CP, Advansys MLP and Advansys DLP systems. Should the Drilling Guides need to be replaced, they can be replenished by the customer placing an order. The replacement is shipped to the customer in sealed plastic bags. Uni-CP" Compression Plate The Uni-CP" Compression Plate is indicated for fixation of bone fractures or for bone reconstruction: - Arthrodesis in foot and ankle surgery - Fracture management in the foot and ankle - Mono or bi-cortical osteotomies in the foot and ankle TIBIAXYS" Plates The Newdeal¿ TIBIAXYS" Plates are indicated For fixation of bone fractures or for bone reconstruction. Examples include Osteotomies and fractures of ankle joint, distal tibia and fibula. The Newdeal¿ TIBIAXYS" Plates must be fixed with the Newdeal Surfix¿ 3.5mm diameter fixation screws and lock-screws. ADVANSYS¿ Mid-Foot Plating System: Dorsal Lisfranc Plate (DLP) and Medical Lisfranc Plate (MLP) The Dorsal Lisfranc Plates are intended for fractures, fusions, osteotomies and replantations of small bones at the tarsometatarsal joints The Medial Lisfranc Plates are intended for bone fixation such as: arthrodesis of the 1st metatarsocuneiform joint to reposition and stabilize a metatarsus primus varus: - Lisfranc arthrodesis - Mono or bi-cortical osteotomies or fractures near the 1st metatarsocuneiform joint

Publicado el 26 de septiembre de 2012 · Inicio del retiro: 14 de junio de 2012

Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Integra' s quality system noted an adverse trend in complaints for breakage of the drilling guides during use. None of the complaints resulted in patient injuries but in some cases resulted in an increase in the time to complete the surgical procedure. A physician was consulted to assess patient risk and concluded that any potential for risk would be transient, self-limiting or minor, or result

Qué hacer

Voluntary: Firm initiated

Detalles del retiro

Fabricante
Integra Life Sci.
Distribuidor
Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MT, NC, NE, NJ, NV, OH, OK, OR, PA, SC, TN, TX, UT, VA, WA, and WI and the countries of Austria, Czech republic, Denmark, France, Germany, Ireland, Israel, Italy, Kingdom of Saudi Arabia, Lebanon, Norway, Portugal, Switzerland, South Africa, Spain, Sweden, and United Kingdom.
Lote
The Lot No's for the affected 2.7mm Drilling Guides are: E1DT, E4EX, E57T, E5YA, E64W, E6GN, E6RB, E6WP, E7UL, E86B, E8UC, E9PX, E9Q5, E9QD, EBNY, EEYJ, EF11, EFCM, EGT6, EH6X, EHL2, EJB3, EJZE, EKJ4, EKN4, EKZV, EKZW, EL3G, EL3X, ELH8, EMCQ, EMCX, EN19, ENR9, ENRA, ENS0, EPTL, and EPTQ
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States