MX 16-slice CT Scanner System, models 728131 and 728132. The MX 16-slice CT Scanner System can be used as a whole body computed tomography X-ray system.
Publicado el 26 de septiembre de 2012 · Inicio del retiro: 24 de agosto de 2012
Resumen
FDA registra el retiro Z-2386-2012 de dispositivos médicos de Clase II, publicado el 26 de septiembre de 2012 e iniciado el 24 de agosto de 2012. Empresa responsable: Philips Medical Systems (Cleveland) Inc; Estado: Finalizado; Cantidad reportada: 328 units; Lote: System Code #, 728131, Serial #: EP16E090001-EP16E090004, EP16E090006-EP16E090015, EP16E100001- EP16E100004, EP16E100006, EP16E100008- EP16E100063, EP16E100067, EP16E100069- EP16E100113, EP16E110001- EP16E110065, EP16E110067-EP16E110070, EP16E110074-EP16E110075, EP16E110077-EP16E110078, EP16E110080, EP16E110083- EP16E110085, EP16E110088- EP16E110112, EP16E110115- EP16E110119, EP16E120001- EP16E120014, EP16E120016, EP16EB110003- EP16ER100007, EP16ER110003, EP16ER110006, P16C080001, P16C080002, P16C090001, P16C100001-P16C100015, P16C100017, P16C100018, P16C110001- P16C110008, P16C110011-P16C110037, P16C120001-P16C120003, P16C120006, P16C120007, P16E080005, P16E080007, P16E080008, P16E080012, P16E090002-P16E090021, P16E090024-P16E090028, P16E100002, P16EF80010, P16EM08003, and P16EM09002-P16EM09004. System Code #, 728132, Serial #: EP16E100005, EP16E100007, EP16E110066, EP16ER100003-EP16ER100005, EP16ER100008, EP16ER110004, EP16ER110005, and EP16ER110007-EP16ER110024. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
On August 24, 2012 the firm recalled the MX 16-slice after discovering that a bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Philips Medical Systems (Cleveland) Inc
- Distribuidor
- Worldwide distribution, Nationwide (US) including the states AL, AR, CA, FL, GA, IL, IN, MS, NJ, NV, NY, OH, PA, and VA and the countries of Albania, Argentina, Australia, Belgium, Brazil, Chile, China, Colombia, Dominican Republic, Ecuador, France, Germany, Greece, India, Indonesia, Ireland, Italy, Jordan, Lebanon, Malawi, Mexico, Nepal, Netherlands, Oman, Palestine, Philippines, Poland, Romania, Russia, Saudi Arabia, South Africa, South Korea, Spain, Switzerland, Turkey, Uganda, and Venezuela.
- Lote
- System Code #, 728131, Serial #: EP16E090001-EP16E090004, EP16E090006-EP16E090015, EP16E100001- EP16E100004, EP16E100006, EP16E100008- EP16E100063, EP16E100067, EP16E100069- EP16E100113, EP16E110001- EP16E110065, EP16E110067-EP16E110070, EP16E110074-EP16E110075, EP16E110077-EP16E110078, EP16E110080, EP16E110083- EP16E110085, EP16E110088- EP16E110112, EP16E110115- EP16E110119, EP16E120001- EP16E120014, EP16E120016, EP16EB110003- EP16ER100007, EP16ER110003, EP16ER110006, P16C080001, P16C080002, P16C090001, P16C100001-P16C100015, P16C100017, P16C100018, P16C110001- P16C110008, P16C110011-P16C110037, P16C120001-P16C120003, P16C120006, P16C120007, P16E080005, P16E080007, P16E080008, P16E080012, P16E090002-P16E090021, P16E090024-P16E090028, P16E100002, P16EF80010, P16EM08003, and P16EM09002-P16EM09004. System Code #, 728132, Serial #: EP16E100005, EP16E100007, EP16E110066, EP16ER100003-EP16ER100005, EP16ER100008, EP16ER110004, EP16ER110005, and EP16ER110007-EP16ER110024.
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDAPinnacle3 Radiation Therapy Planning System, Model numbers 870231 and 870237.21 de septiembre de 2022
- FDABrilliance 64 Model # 728231, computed tomography x-ray system14 de agosto de 2019
- FDAIngenuity Core128 Model # 728323, computed tomography x-ray system14 de agosto de 2019
- FDABrightView XCT , Model Number 88248231 de julio de 2019
- FDASterile Distraction Screws of 3 different sizes: 12mm, 14mm and 16mm sold under TSI, 3COR, DB Surgical, and26 de septiembre de 2012
- FDAInstrumentation Laboratory ACL TOP (Base) PN 0000280000 ACL TOP is a bench top, fully automated, random access26 de septiembre de 2012
Preguntas sobre este retiro
¿MX 16-slice CT Scanner System, models 728131 and 728132. The MX 16-slice CT Scanner System can be used as a whole body computed tomography X-ray system. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2386-2012 el 26 de septiembre de 2012. Peligro reportado: On August 24, 2012 the firm recalled the MX 16-slice after discovering that a bolt which attaches the oil accumulator within the gantry may fail, resulting in the oil accumulator detaching from the tube housing assembly within the gantry.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter