HYGENIC Wingless Fiesta Kit, Model no. H02790 - Product Usage: The dental dam forceps are intended to be used in conjunction with the rubber dental dam and the dental dam clamps. The forceps are used to place and remove the dental dam clamps from the rubber dental dam.
Publicado el 1 de julio de 2020 · Inicio del retiro: 3 de junio de 2020
Resumen
FDA registra el retiro Z-2422-2020 de dispositivos médicos de Clase II, publicado el 1 de julio de 2020 e iniciado el 3 de junio de 2020. Empresa responsable: Coltene Whaledent Inc; Estado: En curso; Cantidad reportada: 7,206 forceps; Lote: Batch numbers: J74546 J74184 If a distributor or end user received one of the batches listed above, they need to look at the batch number laser marked on the forceps themselves. If the forceps has any of the following batch numbers laser marked on them and the forceps do not fit with the dental dam clamps, the forceps should be returned to Coltene. Laser marked batch number: 1903 1904 1906 1907 1909 1910 1911 1915 1920 1921 1923 1927 1928 1930 1934 1940. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Multiple Coltene batches of the product have a pin size larger than acceptable, which may make it difficult for the forceps to fit appropriately with dental dam clamps.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Coltene Whaledent Inc
- Distribuidor
- Pending
- Lote
- Batch numbers: J74546 J74184 If a distributor or end user received one of the batches listed above, they need to look at the batch number laser marked on the forceps themselves. If the forceps has any of the following batch numbers laser marked on them and the forceps do not fit with the dental dam clamps, the forceps should be returned to Coltene. Laser marked batch number: 1903 1904 1906 1907 1909 1910 1911 1915 1920 1921 1923 1927 1928 1930 1934 1940
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACEI Cutting Edge Instruments (PTC-57C)25 de noviembre de 2020
- FDAFG 57 100057X (100057X)25 de noviembre de 2020
- FDAHYGENIC Winged Fiesta Kit, Model no. H02778 - Product Usage: The dental dam forceps are intended to be used in1 de julio de 2020
- FDAAlly Bone Screw - Product Usage:Bone Screws are indicated for use in bone reconstruction, osteotomy,1 de julio de 2020
- FDAStumptown Cold Brew Coffee and Chocolate with Oatmilk. Product is TetraPak Prisma Pak aseptic packaging. Keep1 de julio de 2020
- FDALatex Dental Dams labeled as Non-Latex Dental Dams13 de agosto de 2014
Preguntas sobre este retiro
¿HYGENIC Wingless Fiesta Kit, Model no. H02790 - Product Usage: The dental dam forceps are intended to be used in conjunction with the rubber dental dam and the dental dam clamps. The forceps are used to place and remove the dental dam clamps from the rubber dental dam. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2422-2020 el 1 de julio de 2020. Peligro reportado: Multiple Coltene batches of the product have a pin size larger than acceptable, which may make it difficult for the forceps to fit appropriately with dental dam clamps.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter