STERIS 5085 and 5085 SRT Surgical Tables, Product Usage: The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation. It accommodates all general surgical procedures including but not limited to, cardiac and vascular, endoscopic, gynecology, urology, nephrectomy, neurology, ophthalmologic, orthopedics and other procedures requiring intraoperative fluoroscopic C-arm imaging and also supports laparoscopic surgical technique for the largest surgical patients. The 5085SRT table is a variation of the 5085 table that enables patient transport on hard level surfaces within the surgical suite (from pre-operative areas to the operating room and again from the operating room to post operative recovery).
Publicado el 26 de septiembre de 2012 · Inicio del retiro: 6 de abril de 2012
Resumen
FDA registra el retiro Z-2426-2012 de dispositivos médicos de Clase II, publicado el 26 de septiembre de 2012 e iniciado el 6 de abril de 2012. Empresa responsable: Steris Corporation; Estado: Finalizado; Cantidad reportada: 379 tables; Lote: Serial numbers from 0403809001 to 0408612102. La Clase II indica posibles consecuencias temporales o médicamente reversibles, o una probabilidad remota de consecuencias graves.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
STERIS has learned from Customer feedback and field service experience that damage from external impact to the table hydraulic column can cause the stainless steel shroud sections covering the table hydraulic column to become misaligned; if the damage is severe, the up/down movement of the table may be impeded. In addition, the table hand control may indicate that the table floor locks are unlock
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Steris Corporation
- Distribuidor
- Worldwide Distribution - US Nationwide including the states of: AL, AR, CA, CO, CT, FL, HI, ID, IL, IN, IA, KY, LA, MA, MD, ME, MI, MN, MO, MS, NV, NYH, NC, ND, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, WI & WV and countries of: Australia, Canada, China, India, Korea, Mexico, Panama and Thailand.
- Lote
- Serial numbers from 0403809001 to 0408612102
- Clasificación de riesgo
- Class II
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML02011721 de octubre de 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML02011521 de octubre de 2026
- FDARAPICIDE PA Part A High-Level Disinfectant, Model/Catalog Number: ML02011621 de octubre de 2026
- FDAVERIFY Biological Steam Test Pack Model/Catalog Number: S306914 de octubre de 2026
- FDAULTRA Fabric Reinforced Surgical Gown, X-Large, Sterile, Test File. The Kimberly-Clark ULTRA Fabric Reinforced26 de septiembre de 2012
- FDA12/14 TAPER HIGH OFFSET NECK USE WITH SMITH & NEPHEW STEMS AND 12/14 FEMORAL COMPONENTS ONLY CO-CR, REF26 de septiembre de 2012
Preguntas sobre este retiro
¿STERIS 5085 and 5085 SRT Surgical Tables, Product Usage: The 5085 is a general surgical table with high patient weight capacity, extended width capability, and lower minimal table top elevation. It accommodates all general surgical procedures including but not limited to, cardiac and vascular, endoscopic, gynecology, urology, nephrectomy, neurology, ophthalmologic, orthopedics and other procedures requiring intraoperative fluoroscopic C-arm imaging and also supports laparoscopic surgical technique for the largest surgical patients. The 5085SRT table is a variation of the 5085 table that enables patient transport on hard level surfaces within the surgical suite (from pre-operative areas to the operating room and again from the operating room to post operative recovery). fue retirado del mercado?
Sí. FDA publicó el retiro Z-2426-2012 el 26 de septiembre de 2012. Peligro reportado: STERIS has learned from Customer feedback and field service experience that damage from external impact to the table hydraulic column can cause the stainless steel shroud sections covering the table hydraulic column to become misaligned; if the damage is severe, the up/down movement of the table may be impeded. In addition, the table hand control may indicate that the table floor locks are unlock
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter