Minor Pack 900-2511, pack contains: (1) GOWN XLGE SMS IMPERVIOUS REINFORCED AAMIIII (10) GAUZE SPONGE 4" X 4" 16PLY XRD (5) LAP SPONGE PREWASH 18" X 18" XRD (1) DRAPE LAPAROTOMY W/POUCH 100" x 72" x 124" (4) DRAPE UTILITY WITH TAPE (2) DRAPE SHEET 41" X 69" MEDIUM ( 1) MAYO STAND COVER REINFORCED ( 1) GOWN SURG XLGE REINFORCED TOWEL & WRAP AAMIIII ( 1) BLADE SURGICAL STAINLESS STEEL# 10 (2) LIGHT SHIELD ( 1) CAUTERY PENCIL ROCKER SWITCH ( 1) TABLE COVER REINFORCED 50" X 90" ( 1) UTILITY BOWL BLUE 16oz ( 1) TOWELS ABSORBENT 15" X 20" ( 1) SOLUTION SURGICAL DURAPREP 6ML Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.
Publicado el 10 de septiembre de 2014 · Inicio del retiro: 20 de mayo de 2014
Resumen
FDA registra el retiro Z-2475-2014 de dispositivos médicos de Clase I, publicado el 10 de septiembre de 2014 e iniciado el 20 de mayo de 2014. Empresa responsable: Customed, Inc; Estado: Finalizado; Cantidad reportada: 10 lots (2018 packaged units, multiple units per lot); Lote: Product code 900-2511, 10 lots: 112083072 113099519 113109796 131110292 131210659 140111323 140211706 140211936 140412903 140513479. La Clase I indica una probabilidad razonable de consecuencias graves para la salud o muerte.
Estado reportado al momento de publicación: Terminated — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.
Qué peligro representa
Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
Qué hacer
Voluntary: Firm initiated · Letter
Recibe un correo cuando FDA publique un retiro como este
Gratis. Solo te escribimos cuando hay un retiro nuevo que coincide. Te das de baja con un clic.
Detalles del retiro
- Fabricante
- Customed, Inc
- Distribuidor
- Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.
- Lote
- Product code 900-2511, 10 lots: 112083072 113099519 113109796 131110292 131210659 140111323 140211706 140211936 140412903 140513479
- Clasificación de riesgo
- Class I
- Área (FDA)
- device
- País
- United States
Retiros relacionados
- FDACatalog Number: 9002615 EXTREMITY SURGICAL PACK Used by medical staff to perform surgical procedures and20 de abril de 2016
- FDACatalog Number: 9001482 ENT SURGICAL PACK Used by medical staff to perform surgical procedures and patient20 de abril de 2016
- FDACatalog Number: 9003190 SURG GOWN SMS AMMI IV FULL Used by medical staff to perform surgical procedures and20 de abril de 2016
- FDACatalog Number: 900110 TRACHEOSTOMY CARE KIT WITH TRACH TUBE HOLDER FOR ADULT Used by medical staff to perform20 de abril de 2016
- FDAUrology Pack, 900-2510, package contains: (2) TOWELS ABSORBENT 15X 20 UF (1) DRAPE LITHO 11 0" X 63"10 de septiembre de 2014
- FDALaparotomy Pack, 900-2516, pack contains: ( I) TABLE COVER44" X 90" ( I) CAUTERY PENCIL PUSH BOTTOM (4) TOWELS10 de septiembre de 2014
Preguntas sobre este retiro
¿Minor Pack 900-2511, pack contains: (1) GOWN XLGE SMS IMPERVIOUS REINFORCED AAMIIII (10) GAUZE SPONGE 4" X 4" 16PLY XRD (5) LAP SPONGE PREWASH 18" X 18" XRD (1) DRAPE LAPAROTOMY W/POUCH 100" x 72" x 124" (4) DRAPE UTILITY WITH TAPE (2) DRAPE SHEET 41" X 69" MEDIUM ( 1) MAYO STAND COVER REINFORCED ( 1) GOWN SURG XLGE REINFORCED TOWEL & WRAP AAMIIII ( 1) BLADE SURGICAL STAINLESS STEEL# 10 (2) LIGHT SHIELD ( 1) CAUTERY PENCIL ROCKER SWITCH ( 1) TABLE COVER REINFORCED 50" X 90" ( 1) UTILITY BOWL BLUE 16oz ( 1) TOWELS ABSORBENT 15" X 20" ( 1) SOLUTION SURGICAL DURAPREP 6ML Product Usage: EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents. fue retirado del mercado?
Sí. FDA publicó el retiro Z-2475-2014 el 10 de septiembre de 2014. Peligro reportado: Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.
¿Qué debo hacer si tengo este producto?
Voluntary: Firm initiated · Letter