InPen App, CFN: MMT-8060 (iOS Users) and MMT-8061 (Android Users)

Publicado el 10 de septiembre de 2025 · Inicio del retiro: 16 de junio de 2025

Estado reportado al momento de publicación: Ongoing — openFDA actualiza semanalmente y este campo no refleja el estado más reciente. Confirma siempre con la fuente oficial.

Qué peligro representa

Medtronic MiniMed, Inc. is recalling InPen App for iOS and Android users due to software design errors that could lead to a missed short-acting insulin dose reminder and a recommendation to correct a high glucose value. It does not impact insulin delivery, long-acting insulin dose reminders, or CGM alerts, and users can still use the pen itself to calculate a dose, deliver insulin, record the dose date/time, and view CGM data. This issue was identified during internal testing before release in the US but after release to OUS customers, no complaints or MDRs related to this recall have been reported. Use of the affected device may result hyperglycemia by failing to alert the user and delay treatment of diabetes.

Qué hacer

Voluntary: Firm initiated · Letter

Detalles del retiro

Fabricante
Medtronic MiniMed, Inc.
Distribuidor
International distribution to the countries of Argentina, Australia, Austria, Belgium, Bermuda, Chile, Czech Republic, Denmark, Egypt, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Ireland, Israel, Italy, Luxembourg, Netherlands, New Zealand, Norway, Poland, Portugal, Romania, Saudi Arabia, Serbia, Singapore, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Lote
UDI-DI: 0763000B000122075 CFN/Software Version: MMT-8060/7.0.0, 7.1.0, 7.2.0 and 7.5.0 MMT-8061/7.0.0, 7.0.1, 7.1.0, 7.1.1, 7.2.0 and 7.5.0
Clasificación de riesgo
Class II
Área (FDA)
device
País
United States